ENGLISH OUS indications
Page 46 of 50
OTH-SEM-IFU-OUS-0359 Rev I
Specifications
Table 16 Specifications
Item
Specifications
Applied Part
Type BF
Battery Life
3 hours (typical)
Method of Cleaning and
Disinfection
Clean the device as defined in Section 7
Water ingress protection
IPX1
Duty Cycle
Continuous operation
Power Source
Internally powered equipment
SEM Value Range
1.0 to 4.5 SEM Value units
SEM Value Repeatability
*
1
+/- 0.2 SEM Value units
Storage
The Provizio SEM Scanner should be stored only at
temperatures ranging from -4ºF (-20ºC) to 113ºF (45ºC) at
5% to 90% relative humidity (non-condensing).
Operating Conditions
The Provizio SEM Scanner should be operated only at
temperatures ranging from 59ºF (15ºC) to 95ºF (35ºC) at
5% to 90% relative humidity (non-condensing).
Charging Mat AC Mains Voltage
100-240 V
Charging Mat Mains Current
0.5 A
1
Repeatability is the variation of a single device in measuring the same location by the same person when multiple
measurements are taken one after the other (“repeat”) at that time
.