9
Chapter 3
Specifications
3.1
Compliance with Standards
The CE marking is the manufacturer's declaration that the product meets the requirements of the
applicable EU directives
RCM mark indicates compliance with applicable ACMA technical standards – that is, for
telecommunications, radio communications, EMC and EME
China RoHS mark indicates compliance with administrative measures on the control of pollution caused
by electronic information products according to the Ministry of Information Industries of the People’s
Republic of China
WEEE mark indicates compliance with the EU WEEE Directive
Safety
EN/IEC 61010–1: Safety requirements for electrical equipment for measurement, control and laboratory
use.
UL 61010B–1: Standard for Safety – Electrical measuring and test equipment.
EMC Emission
EN/IEC 61000–6–3: Generic emission standard for residential, commercial and light industrial
environments.
EN/IEC 61000–6–4: Generic emission standard for industrial environments.
CISPR 22: Radio disturbance characteristics of information technology equipment. Class B Limits.
FCC Rules, Part 15: Complies with the limits for a Class B digital device.
EMC Immunity
EN/IEC61000–6–1: Generic standards – Immunity for residential, commercial and light industrial
environments.
EN/IEC 61000–6–2: Generic standards – Immunity for industrial environments.
EN/IEC 61326: Electrical equipment for measurement, control and laboratory use – EMC requirements.
Note:
The above is only guaranteed using accessories listed in the Product Data sheet for Type 4294
and 4294-002 (BP 2101).
Temperature
IEC 60068 –2–1 & IEC 60068–2–2: Environmental Testing. Cold and Dry Heat.
Operating Temperature: 10 to 40
°
C (50 to 104
°
F) for 10 ms
–2
reference within ± 3% and
3.16 ms
–2
reference within ± 3%
–10 to +55
°
C (14 to 131
°
F) for 10 ms
–2
reference within ± 5% and 3.16 ms
–2
reference within ± 5%
Storage Temperature: –25 to +70
°
C (–13 to +158
°
F)
IEC 60068–2–14: Change of temperature: –10 to +55
°
C (2 cycles, 1
°
C/min.)
Humidity
IEC 60068–2–78: Damp Heat: 90% RH (non-condensing at 30
°
C (86
°
F))
Mechanical
Non-operating:
IEC 60068–2–6: Vibration: 0.3 mm, 20 ms
–2
, 10–500 Hz
IEC 60068–2–27: Shock: 1000 ms
–2
IEC 60068–2–29: Bump: 1000 bumps at 400 ms
–2
Enclosure
IEC 60529: Protection provided by enclosures: IP 54
Summary of Contents for 4294
Page 2: ......
Page 6: ......
Page 14: ...Calibration Exciter Types 4294 and 4294 002 User Manual 8...
Page 18: ...Calibration Exciter Types 4294 and 4294 002 User Manual 12...
Page 20: ...Calibration Exciter Types 4294 and 4294 002 User Manual 14...
Page 21: ......