Brooks SLA5810/20 Installation And Operation Manual Download Page 31

3-5

Section 3 Operation

Models SLA5810/20 and SLAMf10/20

Installation and Operation Manual

X-PR-SLA5800-SLAMf-Series-RevB-PC-eng
Part Number: 541B198AAG
January, 2017

RS485 equipped units support the following baud rates. Please specify the
desired baud rate when ordering (default is 19200 baud). Alternately, baud
rate may be changed using the Brooks Expert Support Tool (BEST).

Baud Rates: 1200, 2400, 4800, 9600, 

19200

 and 38400

Reference the Brooks document "Supplemental Manual for RS485
Communications for Brooks

®

 MFCs/MFMs, for SLA5800/SLAMf Revsion B

Series" (X-DPT-RS485-SLA5800-SLAMf-Series-RevB-MFC-PC-RT-eng) for
more details regarding the capabilities of this communication interface.

3-5-2 Profibus Communications Features

The Brooks SLA5810/20 and SLAMF10/20 is now equipped to support the
Profibus communication protocol.

The communication electronics allows for automatic baud rate detection
ranging from 9600 baud to 12 Mbaud, thus making the need for any hardware
baud rate selection methods not required. For selecting the device address,
which must be unique on the network, two rotary switches are provided. This
allows a user to easily select any address number ranging from 0 to 126. This
can provide fast device replacement without complex network configurations.
The device is equipped with a zero command pushbutton, allowing the user to
give a manual command to the device to (re)balance the flow sensor
electronics. This command can also be issued through the protocol.

The Profibus-DP communication option supports the following message types:
• Cyclic data exchange (Write/Read data).
• Read inputs (e.g. status, flow, temperature, totalizer, etc.).
• Read outputs (e.g. commands, setpoint).
• Global control commands (e.g. fail safe, sync).
• Get configuration (i.e. read number of I/O bytes and composition).
• Read diagnostics information (i.e. get error and alarm status).
• Set parameters (i.e. select gas number, engineering units, I/O configuration
• Set parameters (i.e. select gas number, engineering units, I/O
    configuration etc.).
• Check configuration (i.e. check I/O composition).

3-5-3 DeviceNet Communications Features

The Brooks Digital Pressure Controllers are also available with DeviceNet

TM

communication capability. DeviceNet is an open digital protocol capable of
high speeds and easy system connectivity. Brooks Instrument has several of
its devices available on this popular networking standard, and is a member of
ODVA

TM 

(Open DeviceNet Vendors Association), the governing standard body

for DeviceNet.
DeviceNet

 

is similar to the RS485 standard in that it is a multi-drop connection

that allows a maximum of 64 devices to be connected on the same network.
Baud rate selections for DeviceNet

 

products are 125K, 250K and 500K and

can be selected via MAC ID switches mounted on the device.

Summary of Contents for SLA5810/20

Page 1: ...ual X PR SLA5800 SLAMf Series RevB PC eng Part Number 541B198AAG January 2017 Models SLA5810 20 and SLAMf10 20 Brooks Digital Elastomer Sealed Pressure Controllers Models SLA5810 20 and SLAMf10 20 Mod...

Page 2: ...applicable local and national codes Connect all products to the proper electrical and pressure sources Operation 1 Slowly initiate flow into the system Open process valves slowly to avoid flow surges...

Page 3: ...l cell research and chemicals to medical devices analytical instrumentation semiconductor manufacturing and more The Brooks product you have just received is of the highest quality available offering...

Page 4: ...Installation and Operation Manual X PR SLA5800 SLAMf Series RevB PC eng Part Number 541B198AAG January 2017 Models SLA5810 20 and SLAMf10 20 THIS PAGE WAS INTENTIONALLY LEFT BLANK...

Page 5: ...2 6 Removal from Storage 2 2 2 7 Gas Connections 2 2 2 8 In Line Filter 2 2 2 9 Installation 2 3 2 10 Electrical Interface 2 4 Section 3 Operation 3 1 Overview 3 1 3 2 Theory of Operation for Pressur...

Page 6: ...SLAMf10 20 Thru Flow Profibus 1 12 1 10 Dimensions Models SLAMf10 20 Thru Flow DeviceNet 1 12 2 1 Shielded Cable Hookup Diagram Analog RS485 2 5 2 2 Shielded Cable Hookup Diagram Profibus 2 5 2 3 Shie...

Page 7: ...his manual be read in its entirety before attempting to operate or repair these Digital products 1 3 Description Brooks Instrument s digital elastomer sealed pressure measurement and control product l...

Page 8: ...can directly replace existing analog pressure controllers with the advantage of improved digital accuracy Both analog and digital applications will see settling time and accuracy improvements 1 4 Spec...

Page 9: ...psia 20 7 bara for 100 300 psia full scale 300 psig 20 7 barg for 100 300 psig full scale 3000 psia 206 9 bara for 300 3000 psia full scale 4500 psia 310 3 bara for 3000 4500 psia full scale Leak Int...

Page 10: ...Flow Output Voltage Specifications Flow Output Voltage Specifications Flow Output Voltage Specifications Flow Output Voltage Specifications Flow Output Voltage Specifications Nominal Range Nominal Ra...

Page 11: ...EU Directive 2011 65 EU EN 61326 1 2013 EMC RoHS ATEX IECEx Special Conditions for safe use 1 The module shall be installed in a suitable enclosure providing a degree of protection of at least IP54 ac...

Page 12: ...temperature specification of the selected cable shall be in compliance with the actual measured temperature values 2 Provisions shall be made to prevent the rated voltage from being exceeded by transi...

Page 13: ...d rate selections for the Brooks Digital Series related to RS485 are 1200 2400 4800 9600 19200 and 38400 baud and can be selected via the Brooks Expert Support Tool BEST The RS485 is essentially a mul...

Page 14: ...k connection The communication electronics allows for automatic baud rate detection ranging from 9600 baud to 12 Mbaud thus making the need for any hardware baud rate selection methods not required Fo...

Page 15: ...is similar to the RS485 standard in that it is a multi drop connection that allows a maximum of 64 devices to be connected on the same network Baud rate selections for DeviceNetproducts are 125K 250K...

Page 16: ...Series RevB PC eng Part Number 541B198AAG January 2017 Figure 1 6 Dimensions Models SLA5810 20 Thru Flow Profibus Figure 1 5 Dimensions Models SLA5810 20 Thru Flow RS485 Note Aux Input is used for Re...

Page 17: ...Series RevB PC eng Part Number 541B198AAG January 2017 Figure 1 8 Dimensions Models SLAMf10 20 Thru Flow RS485 Figure 1 7 Dimensions Models SLA5810 20 Thru Flow DeviceNet Note Aux Input is used for R...

Page 18: ...eries RevB PC eng Part Number 541B198AAG January 2017 Figure 1 10 Dimensions Models SLAMf10 20 Thru Flow DeviceNet Figure 1 9 Dimensions Models SLAMf10 20 Thru Flow Profibus Note Aux Input is used for...

Page 19: ...ortCenters 2 3 Recommended Storage Practice If intermediate or long term storage of equipment is required it is recommended that the equipment be stored in accordance with the following a Within the o...

Page 20: ...ess of those recommended See Section 2 3 the device should be subjected to a pneumatic pressure test in accordance with applicable vessel codes 2 7 Gas Connections Prior to installation ensure all pip...

Page 21: ...ontroller should be located in a clean dry atmosphere relatively free from shock and vibration b Install in such a manner that permits easy removal if the instrument requires servicing c Leave suffici...

Page 22: ...nal Common Signal Output 13 5 to 27 Vdc Supply Setpoint Input Setpoint Common Supply Common For a DeviceNet unit 11 25 Vdc power and communication I O are supplied via the standard 5 pin Circular Micr...

Page 23: ...0 SLAMF Series RevB PC eng Part Number 541B198AAG January 2017 Figure 2 2 Shielded Cable Hookup Diagram Profibus Figure 2 1 Shielded Cable Hookup Diagram Analog RS485 Note Aux Input only used for Remo...

Page 24: ...lded Cable Hookup Diagram DeviceNet Notes 1 The commons for the setpoint and signal output are not isolated from power supply common 2 The cable shields should connect to chassis ground at one end onl...

Page 25: ...Operation Manual X PR SLA5800 SLAMF Series RevB PC eng Part Number 541B198AAG January 2017 Figure 2 5 Recommended I O Wiring Configuration for Current Signals Non Isolated Power Supply Figure 2 6 Rec...

Page 26: ...Section 2 Installation Models SLA5810 20 and SLAMf10 20 Installation and Operation Manual X PR SLA5800 SLAMf Series RevB PC eng Part Number 541B198AAG January 2017 THIS PAGE WAS INTENTIONALLY LEFT BL...

Page 27: ...transducer control electronics and control valve into one unit results in a compact size The mounting dimensions are comparable to a mass flow controller Refer to Figures 1 5 to 1 10 The upstream pre...

Page 28: ...s pressures full scale flow rates The supplemental manuals for Profibus DeviceNet and RS485 contain additional details regarding specific communication features Calibrations will appear in the calibra...

Page 29: ...clude any of the following I O options as specified by the user 0 5 Vdc setpoint 0 5 Vdc flow output 1 5 Vdc setpoint 1 5 Vdc flow output 0 10 Vdc setpoint 0 10 Vdc flow output 0 20 mA setpoint 0 20 m...

Page 30: ...been zeroed at the factory to a known vacuum condition It should not require any further adjustment Reference Section 2 9 for further details 3 5 Communications Features 3 5 1 Analog RS485 Communicat...

Page 31: ...vice is equipped with a zero command pushbutton allowing the user to give a manual command to the device to re balance the flow sensor electronics This command can also be issued through the protocol...

Page 32: ...The DeviceNetcommunication link also provides access to many of the Brooks SLAMf Digital Series functions for control and monitor operations including Accurate setpoint adjustment and flow output mea...

Page 33: ...ssary flows then the response of the pressure controller may be sluggish or slow For instance a command to an upstream pressure controller to increase pressure in a vessel will cause the valve to clos...

Page 34: ...lly if the system is new 1 Verify a low resistance common connection and that the correct power supply voltage and signals are reaching and leaving the controller The Brooks Expert Support Tool will m...

Page 35: ...tion and Operation Manual X PR SLA5800 SLAMf Series RevB PC eng Part Number 541B198AAG January 2017 Models SLA5810 20 and SLAMf10 20 Section 4 Maintenance Troubleshooting Figure 4 1 Bench Troubleshoot...

Page 36: ...ation and Operation Manual X PR SLA5800 SLAMf Series RevB PC eng Part Number 541B198AAG January 2017 Models SLA5810 20 and SLAMf10 20 Section 4 Maintenance Troubleshooting THIS PAGE WAS INTENTIONALLY...

Page 37: ...ls SLA5810 20 and SLAMf10 20 Brooks Instrument Brooks Instrument x x x Brooks Instrument x x x 1 2 x Brooks Instrument x PED 0 5 bar g 25 mm 1 inch PED x PED x 2014 34 EU x Brooks Instrument 1 x 25 mm...

Page 38: ...je nutn za zen kompletn vypr zdnit Pokud tak neu in te m e z d vodu teplotn rozta nosti zbytk m dia v za zen doj t k jeho po kozen nebo k ohro en zdrav osob Evropsk sm rnice pro tlakov za zen PED Na v...

Page 39: ...ventilen ikke bliver lukket p samme tid i forbindelse med aftapning S fremt dette ikke overholdes er der risiko for at v sken p grund af varmeudvikling ekspanderer og dette kan for rsage skade p udsty...

Page 40: ...en afgetapt Als dat wordt nagelaten kan dit leiden tot thermische expansie van de vloeistof waardoor het apparaat kan barsten en lichamelijk letsel kan veroorzaken PED Pressure Equipment Directive All...

Page 41: ...sta Vastasel korral v ib vedelik soojuse m jul paisuda niiv rd et seade puruneb See v ib p hjustada raskeid kehavigastusi Euroopa surveseadmete direktiiv PED Euroopa surveseadmete direktiiv kohaldub k...

Page 42: ...nti aiheuttaa nesteen l mp laajenemista joka saattaa johtaa laitteen rikkoutumiseen ja henkil vahingon vaaraan Eurooppalainen painelaitedirektiivi PED Painelaitteet joiden paine on suurempi kuin 0 5 b...

Page 43: ...ment vidang s Si cela n est pas fait une ventuelle dilatation thermique du fluide peut casser l appareil et provoquer des blessures Directive europ enne quipements sous pression PED Tous les quipement...

Page 44: ...ventilen komplett entleert werden anderenfalls kann es zu einer thermischen Ausdehnung der Fl ssigkeit und damit zum Bruch des Ger tes kommen Personensch den k nnen die Folge sein Europ ische Druckger...

Page 45: ...y 2017 Models SLA5810 20 and SLAMf10 20 Greek Brooks Instrument Brooks Instrument x x x Brooks Instrument x x x 1 2 x Brooks Instrument x PED 0 5 bar g 25 mm 1 PED x PED x 2014 34 EU x Brooks Instrume...

Page 46: ...i ki s bel p szelepeket a berendez st teljesen le kell r teni Ennek elmulaszt sa a folyad k h t gul s t okozhatja ami k ros thatja a berendez st s szem lyi s r l shez vezethet Nyom start berendez sekr...

Page 47: ...lora si deve svuotare completamente l apparecchiatura L inosservanza pu causare la dilatazione termica del liquido che pu danneggiare l apparecchiatura e provocare lesioni alle persone Direttiva europ...

Page 48: ...dr s pl smas ier u tuvum eso os iepl des un izpl des v rstus k da iemesla d ir j aizver no ier c m ir j izlai viss idrums Pret j gad jum idrums var termiski izplesties p rraut ier ci un rad t traumas...

Page 49: ...idimo vo tuvus i renginio reikia i leisti vis skyst To nepadarius galimas iluminis skys io pl timasis galintis sugadinti rengin ir su eisti mon s Europos sl gin s rangos direktyva PED Visa sl gin rang...

Page 50: ...cieczy je eli znajduj ce si na nich zawory wej ciowe i wyj ciowe maj by z jakiego powodu zamkni te to urz dzenie musi zosta ca kowicie opr nione z ciek ego medium Niedope nienie tego zalecenia mo e do...

Page 51: ...rio fechar as v lvulas a montante e jusante do equipamento tenha em aten o que o mesmo dever ser previamente esvaziado do flu do que o atravessa Esta obriga o surge do facto de existir a possibilidade...

Page 52: ...iderea valvelor de intrare i ie ire limitrofe instala iei instala ia trebuie complet golit Neglijarea acestui lucru poate avea ca efect dilatarea termic a fluidului care poate defecta instala ia i poa...

Page 53: ...hoko vek d vodu je nutn uzavrie vstupn a v stupn ventily zariadenie je potrebn plne vypr zdni Zanedbanie vyp ania m za n sledok po kodenie zariadenia s mo nos ou zranenia obsluhy z d vodu tepelnej ro...

Page 54: ...popolnoma izprazniti e tega ne naredite lahko pride do toplotnega raztezanja teko ine zaradi katerega se lahko naprava prelomi in povrzo i telesne po kodbe Evropska direktiva o tla ni opreni PED Vsa t...

Page 55: ...lado del equipo dichas instalaciones deber n ser completamente vaciadas La omisi n de esto ltimo puede provocar la dilataci n t rmica del l quido lo que puede da ar al equipo y conducir a da os perso...

Page 56: ...skada utrustningen och orsaka personliga skador European Pressure Equipment Directive PED R dets Direktiv 99 36 EG av den 29 april 1999 1 om transportabla tryckb rande anordningar Alla utrustning f r...

Page 57: ...Installation and Operation Manual X PR SLA5800 SLAMf Series RevB PC eng Part Number 541B198AAG January 2017 Models SLA5810 20 and SLAMf10 20 THIS PAGE WAS INTENTIONALLY LEFT BLANK...

Page 58: ...rds Visit www BrooksInstrument com to locate the service location nearest to you START UP SERVICE AND IN SITU CALIBRATION Brooks Instrument can provide start up service prior to operation when require...

Reviews: