Installation and Operation Manual
X-CM-QmB-eng
Part Number: 541B029AAG
June, 2017
QmB Series IP40, IP66, IP66XP
Figure 2-3 Horizontal Right Side Up Installation
topside of the pipe to limit the possibility of entrapped fluid collecting in the
sensor, which can cause errors. (See Figure 2-3).
Vertical Mounting
Figure 2-4 Vertical Flow Up Installation
If the Brooks Quantim is installed in a vertical orientation, for use in a
liquid application liquid should flow upwards through the instrument to help
minimize errors due to entrapped gas (See Figure 2-4). For gas flow vertical
orientation, flow up or down is acceptable.
2-5