3. Installation
CryoPod Carrier
CryoPod Carrier Assembly
Part Number: 257437 Rev. A
CryoPod Carrier Assembly
Step
Action
1.
Remove the battery panel on the lower front of the unit using the built-in clip.
2.
Insert the provided 3 "AA" alkaline batteries in the battery wells as indicated by
the embossed schematic. When replacing batteries, all three must be replaced
at a time.
3.
Replace the battery panel.
4.
Put the instrument lid on.
5.
Turn the CryoPod Carrier on by pressing the power button
. The LCD
screen will display the start-up sequence.
6.
Enter SET Mode by long pressing the time button
. The SET text will
display.
7.
Short press
to consecutively cycle through settable elements.
Elements appear in the following order:
l
Alarm 1 (TWARNING) set temperature,
l
Alarm 1 ON/OFF,
l
Alarm 1 sound ON/OFF,
l
Alarm 2 (TCRITICAL) set temperature,
l
Alarm 2 ON/OFF,
l
Alarm 2 sound ON/OFF,
l
Date (day), Date (Month), Date (Year),
l
Time (hr), Time (min), Time (AM/PM),
l
ID number,
l
Standby (TSTANDBY) set temperature
8.
Set values for each settable element by short pressing the star burst button
or the silent alarm button
. Once the value is set, short press
to move to the next functional element.
21