Section 2 Installation
Position the SolidSense II pressure transducer with the electrical
connector facing pointing upward. Using a small, flat-bladed screw driver,
turn zero adjustment screw in a clockwise direction to increase the signal.
Turn the screw in a counterclockwise direction to decrease the signal.
Figure 2-6 Location of the Zero Adjustment Screw on the Standard
SolidSense II Pressure Tranducers
Red (+) Excitation
Black (-) Common
(Bare) Drain Wire
Zero Adjustment Screw
Zero Adjustment
The SolidSense II pressure transducer is factory calibrated and does not
normally need a field adjustment. If a field adjustment becomes necessary,
adjust the zero as noted below.
Zero Adjustment (SolidSense II Standard Pressure Transducers)
On standard SolidSense II pressure transducer models, the zero
adjustment screw is located on the body as shown in Figure 2-6.
SolidSense II® Series Pressure Transducer
0
7
/2022
23