Section 2 Installation
2-5
Installation and Operation Manual
X-CM-QmB-eng
Part Number: 541B029AHG
June, 2008
Horizontal Mounting
If installing the Brooks QUANTIM in a horizontal orientation, liquids should
flow in the direction of the flow arrow. Situations of entrapped gas in the
liquid process should be avoided. In liquid flow applications, it is
recommended to mount the QUANTIM in an inverted attitude on the
underside of the pipe to limit the possibilitiy of entrapped gas collecting in
the sensor, which can cause errors (See Figure 2-2).
For use in gas flow measurement, the QUANTIM should be mounted on
the topside of the pipe to limit the possibility of entrapped fluid collecting in
the sensor, which can cause errors. (See Figure 2-3).
Figure 2-2 Horizontal Inverted Installation
Figure 2-3 Horizontal Right Side Up Installation