Brookfield Engineering Laboratories, Inc.
Page 45
Manual No. M98-211-E0912
Click the Delete button to delete the command in the selected row of the program
grid.
Click the Up button to move the command in the selected row of the program grid up
one row.
Click the Down button to move the command in the selected row of the program
grid down one row.
Click the Clear button to clear the grid of all commands. Once cleared, the
commands cannot be retrieved.
Figure IV-31
This list box displayed the commands available for
creating programs. As previously stated, clicking on the
Insert button inserts the highlighted command (WTI in this
case) into the selected line in the program grid. Double-
clicking on a line in this list box also inserts the command
into the grid.
The icons to the left of the command descriptions indicate
the type of command:
A command to wait for a condition.
A command to set a program parameter.
A command to send information to an attached
printer.
Before downloading a program to the Rheometer, ensure the following have been checked:
• The appropriate cable (BEL Part No. DVP-80 or DVP-202) is connected between the se-
lected COM port/USB of the PC and the Rheometer.
• The DV-III Ultra is at the download screen: OPTIONS/DOWNLOAD A PROGRAM/
LOAD TO SLOT#x where x is slot 1,2,3, or 4.
• With the LOAD TO SLOT #x screen displayed, choose a store slot using the DV-III Ultra
arrow keys, then press the ENTER/AUTORANGE key on the Rheometer. If after five sec-
onds, the Rheometer cannot communicate with the RheoLoader program, the B.E.V.I.S.