Brookfield Engineering Laboratories, Inc.
Page 22
Manual No. M98-211-E0912
VISCOMETER SETUP
PRESS ENTER TO START
Figure II-9
Press the
ENTER
key. The screen in
Figure
II-10 appears. Press the "3" key for NO.
PRINT PARAM RAM?
1 = YES
3 = NO
Figure II-10
The display will now ask “SPECIAL SPINDLE?” as shown in
Figure
II-11. Press the “1” key
for YES.
SPECIAL SPINDLE?
1 = YES
3 = NO
Figure II-11
The display will ask “ADD A SPINDLE?” Press the “1” key for YES.
ADD A SPINDLE?
1 = YES
3 = NO
Figure II-12
The spindle number and name will be flashing. Use the “2” key for scrolling downward (
ê
) and
the “8” key for scrolling upward (
è
). The spindle name can be changed to any (2) letters from A
to Z.
Use the “tab” key to move left and right through the menu. Press the “enter” key to accept.
SPECIAL SPINDLE #1
NAME = D AA E
SMC =
SRC =
Figure II-13
Check the spindle constants to be entered. You must have the SMC (spindle multiplier constant)
and the SRC (the shear rate constant). Enter the first set of numbers. Press the “enter” key to go
to the next line (example: spindle AA = SMC = 1.0 SRC = 2.0).
If you are adding additional spindles, press the “enter” key. You can enter up to 5 special
spindles.