Bronkhorst®
Instruction Manual MI Series Industrial mini CORI-FLOW™
9.17.120A
35
Reset Counter Enable
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…15
157
104/9
0x0D09/3338
Available reset methods for counters. Up to 3 different methods can be specified. The value is a bit-wise summation of the
enabled reset methods; convert the value to binary to see which methods are enabled.
Default value: 7 (bits/methods 0, 1 and 2 enabled)
The following methods are supported:
Bit
Value
Description
0
1
Micro-switch
1
2
Externally
2
4
By parameter
Reset
3
8
Automatically (e.g. when counter value is reset)
4.4
Network configuration
Changes made to the network settings will
not
be restored by a factory reset.
Default settings
Network configuration is done ex factory as indicated on the serial number label or in the technical specifications. The table
below shows the supported configurations for the available interface protocols (default settings are printed in boldface):
Protocol
ProPar
(RS232)
FLOW-BUS
(RS485)
Modbus
(RTU/ASCII)
PROFIBUS DP
HART
Bus Address
3
3
…125
1
…247
0…
126
0
Baud Rate
9600
19200
38400
57600
115200
230400
460800
187500
400000
9600
19200
38400
56000
57600
115200
128000
256000
(autodetect)
9600
19200
45450
93750
187500
500000
1500000
3000000
6000000
12000000
1200
Bus Parity
0
0
0, 1,
2
2
1
EtherCAT® and PROFINET configuration is done automatically via the Ethernet protocol.
Fieldbus Interface Index
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…4
378
125/7
0x0FA7/4008
This parameter indicates the communication interface to which
Fieldbus 1 Address
,
Fieldbus 1 Baud Rate
and
Fieldbus 1 Parity
apply (see further). Each specific interface has its own index and corresponding network parameter values:
Value
Description
0
Fieldbus (fieldbus terminal B/C)
1
Fieldbus (main terminal M)
2
Display
3
Service port
4
HART