Bronkhorst®
Instruction Manual ES-FLOW™ 1xxC
9.17.153A
19
5.5.1.3
Network settings - readout/change
Reading network settings
·
By briefly pressing the switch 3 times with intervals of up to 1 second in normal operation mode, the instrument shows
its current node address and baud rate with a series of consecutive LED indication patterns:
Step Pattern
Indication
1.
Green
l
l
number of flashes indicates the tens of the node address
2.
Red
l
l
number of flashes indicates the units of the node address
3.
Green and red
(simultaneous)
l
l
number of flashes indicates the baud rate
Examples:
·
for node address 35, the green LED will flash 3 times and the red LED 5 times
·
for node address 116, the green LED will flash 11 times and the red LED 6 times
On DeviceNet™ the node address is called MAC ID.
The number of flashes for the baud rate indication is associated with the following baud rates:
Number of
flashes
(index)
Baud rate
FLOW-BUS
Modbus
(ASCII/RTU)
PROFIBUS DP
CANopen
DeviceNet™
Ethernet based
0
automatically
detected
1
187500
9600
9600
1000000
125000
100000000
2
400000
19200
19200
800000
250000
3
38400
45450
500000
500000
4
56000
93750
250000
5
57600
187500
125000
6
115200
500000
50000
7
128000
1500000
20000
8
256000
3000000
10000
9
6000000
10
12000000