background image

© 2018 Bronkhorst High-Tech B.V.

3

c. Ambient conditions

EL-FLOW® Prestige

 instruments are suitable for use in conditions between -10…+70 

°

C and relative humidity of

10…90 %RH, unless specified otherwise. Please note that 

EL-FLOW® Prestige

 instruments have an ingress

protection rating of IP40, implying that the electronics housing and electrical connection do not offer any
protection against moist environments.

3

Check if system piping is clean

For reliable measurement, always make sure the fluid stream is clean.
Use filters to assure a moisture, oil and particle free gas stream
(recommended pore-size: 5 µm).
If back flow can occur, installing a downstream filter and a check valve is
recommended too. For high flow rates, select a suitable filter size, to avoid
a too high pressure drop.

4

Mount/install instrument properly

Mount the instrument in the line in accordance with the flow arrow in the
instrument body. Tighten fittings according to the instructions of their
supplier. The use of Swagelok RS-type stainless steel adapters is advised.

Avoid installation in close proximity of mechanic vibration and/or heat sources.

5

Leak check

Check the system for leaks before applying (fluid) pressure, especially if toxic, explosive or other dangerous fluids
are used.

6

Electrical connection

Electrical connections must be made with a standard cable or according to
the applicable hook-up diagram (see page 2). The factory installed 9-pin
sub-D settings are indicated on the instrument label. Make sure that the
power supply is suitable for the indicated power ratings, indicated on the
instrument label, and that double or reinforced insulation is used for the
power supply.

7

a. Analog/local operation

Connect the 

EL-FLOW® Prestige

 to the power supply/readout unit using a

cable with 9-pin sub-D connector.

b. Digital RS232 operation

Digital operation over RS232 can be established when using the
following setup. Connecting the instrument with a RS232 cable
or an RS232 cable with a USB to RS232 converter to a PC
enables use of the (free) Bronkhorst® software for Windows,
such as FlowDDE and FlowPlot. Make sure that the instrument
label indicates RS232 settings and apply the proper baud rate
settings. If the instrument is not set for RS232 communication,
please refer to the EL-FLOW® Prestige manual (document no.
9.17.084) for switching to RS232 communication settings.

Reviews: