9.17.083 E

2

SCOPE OF THIS GUIDE

EL-FLOW® Prestige 

is the next generation of Bronkhorst® Mass Flow Meters/ Controllers, using the latest, highly accurate,

thermal bypass sensor technology and featuring excellent control characteristics. Thanks to the advanced sensor technology in
combination with a state-of-the-art printed circuit board, the instruments feature improved long term stability as well as an
unsurpassed temperature stability. The optional Pressure Insensitive functionality offers real-time pressure compensation
ensuring additional accuracy and reliability under varying inlet pressure conditions. Furthermore, due to a power efficient
microprocessor, the power consumption of the instrument is reduced. The mass flow meters and controllers can be supplied in
full scale ranges from 0,7 ml

n

/min up to 100 l

n

/min Air-equivalent at max. 64 or 100 bar (1000 or 1500 psi) pressure rating.

This Quick Installation Guide will help you start up your 

EL-FLOW®

 

Prestige

 in only 10 steps, covering the following subjects:

1.

  Check functional properties

2.

  Check pressure

3. 

 Check piping

4.  

Mounting/installation

5.

  Leak check

6

.   Electrical connection

7.

   Operational interface

8. 

  Multifunctional switch

9. 

  Purging

10.

 Zeroing

Other applicable documents:

Manual for EL-FLOW® Prestige series

9.17.084

FlowPlot manual

9.17.030

Fieldbus/interface manuals

Manual

Hook-up diagram

RS232 interface with FLOW-BUS protocol 9.17.027

9.16.119 (incl. analog interface)

FLOW-BUS interface

9.17.024

9.16.120

PROFIBUS DP interface

9.17.025

9.16.121

DeviceNet™ interface

9.17.026

9.16.122

Modbus interface

9.17.035

9.16.123

PROFINET interface

9.17.095

9.16.147

EtherCAT® interface

9.17.063

9.16.124

Custom bus & IO configuration

9.16.118

Micro switch control pin

With this pin the micro switch 
on top of the EL-FLOW® Prestige 
can be operated.

   

These documents can be downloaded from 

www.bronkhorst.com/downloads

 or can be sent by email on request.

Starting up

1

Check functional properties

Before installing the instrument, check if the properties stated on the
instrument label match your requirements:

Flow rate

Fluid to be measured

Up- and downstream pressures

Input/output signals 

(see also step 6)

Temperature

Valve type 

(N.C. - Normally Closed / N.O. - Normally Opened)

2

a. Check test-pressure

The tested pressure is stated on the instrument with a red sticker.
Before installation, make sure that the tested pressure is in accordance with
normal safety factors for your application. If a pressure test sticker is
missing, or if the tested pressure is insufficient, the instrument must 

not 

be

mounted in the process line and should be returned to the factory.

b. Sealings

EL-FLOW® Prestige

 instruments are equipped from factory with sealings compatible with the requested gas type.

However the instruments have Multi Gas / Multi Range functionality on board. If another gas or mixture is used,
always make sure that the gas/mixture is compatible with the installed sealing materials. Refer to the EL-FLOW®
Prestige manual (document no. 9.17.084) for a compatibility list.

© 2018 Bronkhorst High-Tech B.V.

3

c. Ambient conditions

EL-FLOW® Prestige

 instruments are suitable for use in conditions between -10…+70 

°

C and relative humidity of

10…90 %RH, unless specified otherwise. Please note that 

EL-FLOW® Prestige

 instruments have an ingress

protection rating of IP40, implying that the electronics housing and electrical connection do not offer any
protection against moist environments.

3

Check if system piping is clean

For reliable measurement, always make sure the fluid stream is clean.
Use filters to assure a moisture, oil and particle free gas stream
(recommended pore-size: 5 µm).
If back flow can occur, installing a downstream filter and a check valve is
recommended too. For high flow rates, select a suitable filter size, to avoid
a too high pressure drop.

4

Mount/install instrument properly

Mount the instrument in the line in accordance with the flow arrow in the
instrument body. Tighten fittings according to the instructions of their
supplier. The use of Swagelok RS-type stainless steel adapters is advised.

Avoid installation in close proximity of mechanic vibration and/or heat sources.

5

Leak check

Check the system for leaks before applying (fluid) pressure, especially if toxic, explosive or other dangerous fluids
are used.

6

Electrical connection

Electrical connections must be made with a standard cable or according to
the applicable hook-up diagram (see page 2). The factory installed 9-pin
sub-D settings are indicated on the instrument label. Make sure that the
power supply is suitable for the indicated power ratings, indicated on the
instrument label, and that double or reinforced insulation is used for the
power supply.

7

a. Analog/local operation

Connect the 

EL-FLOW® Prestige

 to the power supply/readout unit using a

cable with 9-pin sub-D connector.

b. Digital RS232 operation

Digital operation over RS232 can be established when using the
following setup. Connecting the instrument with a RS232 cable
or an RS232 cable with a USB to RS232 converter to a PC
enables use of the (free) Bronkhorst® software for Windows,
such as FlowDDE and FlowPlot. Make sure that the instrument
label indicates RS232 settings and apply the proper baud rate
settings. If the instrument is not set for RS232 communication,
please refer to the EL-FLOW® Prestige manual (document no.
9.17.084) for switching to RS232 communication settings.