BRONKHORST CORI-TECH B.V.
2.4 Mounting
Install the CORI-FLOW in accordance to the direction of the FLOW arrow. The arrow for flow direction is
indicated on the instrument, between process fittings.
Mount the CORI-FLOW on a firm and steady base.
Avoid in any case heavy accelerations or mechanical shocks on the instrument.
For liquids, mount the CORI-FLOW at a level in the pipe line system where gas enclosure
is not possible and
be certain that the CORI-FLOW is filled all the time while operating.
For gases mount the CORI-FLOW at a level in the pipe line system where condensate can not collect in the
CORI-FLOW.
Verification of leaks is required prior starting up of the process
.
page 12