Power Supply
• The appliance must not be plugged into a multiple outlet power board.
• The appliance is supplied with a flexible power cord fitted with a 3-pin plug.
• The socket used must be properly earthed and fuse protected.
• See the rating label inside the cabinet for supply voltage, frequency and current draw.
• Power cable is rated 70°C. It is recommended that the cable should not come into contact
with parts over 50°C.
• Do not turn o
ff
power before consulting the person responsible.
• Mark the power source clearly, so the refrigerator is not unplugged or turned o
ff
accidentally.
Before Operating
• If the appliance has been sharply tilted or transported in the horizontal position, it must
be left to stand in the upright position and not plugged in for 1 hour.
• Remove all the packaging material before using the appliance.
• Start the appliance only if it has been installed as described.
1300 276 642
www.bromicrefrigeration.com.au
6
NOTE:
The maximum ambient temperature for the refrigerator is 30°C. Conditions
outside this temperature will result in cabinet temperatures outside the required 2
to 8°C range. Bromic recommends that the refrigerator is installed at an ambient
temperature of 20 to 25°C.
4. INSTALLATION
Locating the Cabinet
Ensure the following instructions are observed as this is vital for proper and trouble-free
operation.
• Do not install the cabinet outdoors.
• Locate the appliance in a dry and well ventilated place.
• The unit must be positioned on a level surface.
• Do not install in the vicinity of heating radiators or other heat sources.
• Do not install the appliance where it is exposed to direct sunlight.
• Do not install the appliance where it is exposed to warm draughts or open doors.
• Keep the air intake and outlets free of obstructions.
• Adequate allowance should be made for door openings.
Ventilation
Ensure the following instructions are observed as this is vital for proper and trouble-free
operation.
• Allow 100mm clearance between walls and the appliance.
• Ensure that the condenser fan is free to take in ambient air from the room.
• Adherence to minimum clearances is necessary
SIDE VIEW
100mm
100mm
100mm
100mm
FRONT VIEW