MJ1023 GB Issue 01
7
5.3
HIGH TEMPERATURE ALARM DISPLAY
– If the measured temperature goes up by more than the
figure in the ALARM HI screen, the
alarm will sound immediately and “+T” will be displayed. Press OK
to silence the alarm for 30 minutes.
If the high temperature problem rectifies itself the “+T” remains on the display to show this has
happened. Press OK to clear the indicator. Check the incubator is not (and has not been) in direct
sunlight or too near a heat source such as a room heater. It is advisable to candle eggs a number of
times after this event to check for losses.
5.4
LOW TEMPERATURE ALARM DISPLAY
– If the measured temperature goes down by more than the
figure in the ALARM LO screen, after 6
0 minutes “-T” will be displayed and the alarm will sound. Press
OK to silence the alarm for 30 minutes.
If the low temperature problem rectifies itself the “-T” remains on the display to show this has
happened. Press OK to clear the indicator. Check the incubator is not (and has not been) in a cold
draught or that the room temperature has dropped significantly. It is advisable to candle eggs a
number of times after this event to check for losses.
5.5
LOW ROOM TEMPERATURE ALARM
– If the calculated room temperature remains too low for
optimum results for more than 1 hour a warning is displayed
“-RM” and an alarm will sound. Press OK
to silence the alarm for 30 minutes.
If the low temperature
problem rectifies itself the “-RM” remains on the display to show this has
happened. Press OK to clear the indicator. Check the incubator is not (and has not been) in a cold
draught or that the room temperature has dropped significantly. It is advisable to candle eggs a
number of times after this event to check for losses.
5.6 HIGH ROOM TEMPERATURE ALARM
– If the calculated room temperature remains too high for
optimum results for more than 1 hour a warning is displayed
“+RM” and an alarm will sound. Press OK
to silence the alarm for 30 minutes.
If the high temperature
problem rectifies itself the “+RM” remains on the display to show this has
happened. Press OK to clear the indicator. Check the incubator is not (and has not been) in direct
sunlight or too near a heat source such as a room heater. The eggs themselves create significant
metabolic heating at later stages of incubation and may contribute to this if room temperature is high it
is advisable to candle eggs a number of times after this event to check for losses.
T 39.8C
H 45%
+
T/
T*32.1C
H 45%
-
T/
T*37.5C
-RM /
T*37.5C
+RM /