MJ1023 GB Issue 01
18
CALIBRATION MENU
CAL TEMP
OK
CAL RH%
OK
RH CYCLE
OK
AMBIENT
OK
SAVE
OK
CANCEL
OK
T 37.3C
OK CAL!
RH 48%
OK CAL!
RHC 7S
OK
-
+
T 20.0C
OK
-
+
CALIBRATE THERMOMETER.
TAKE READINGS AT 30mm ABOVE THE
CENTRE OF EACH EGG TRAY AND
CALCULATE THE AVERAGE.
CALIBRATE HYGROMETER.
PLACE HYGROMETER IN CENTRE WITH
NO WATER IN THE EVAPORATING TRAY.
HUMIDITY PUMP CYCLE TIME.
ONLY APPLICABLE WHEN USED WITH
THE BRINSEA ADVANCE HUMIDITY
PUMP.
AMBIENT TEMPERATURE
COMPENSATION.
AMBIENT TEMPERATURE USED IN
HEATER CONTROL ALGORITHM. ONLY
ADJUST IF DISPLAY VARIES BY MORE
THAN +/- 0.2°C.
TURN ALM
OK
T ALM:1
OK 0 1
TURN ALARM.
SWITCHES THE TURN WARNING ALARM
OFF IF CAUSING A DISTURBANCE. IT IS
RECOMMENDED TO LEAVE THIS
SWITCHED ON.
-RM %
OK
50%
OK
-
+
MAINS FREQUENCY SELECTION
(Region Specific)
SWITCHES BETWEEN
50Hz (EUROPE, etc.)
& 60Hz (NORTH AMERICA, JAPAN, etc)
+RM %
OK
5%
OK
-
+
HZ
OK
HZ
OK 50 60
LOW ROOM TEMPERATURE ALARM
ADJUSTMENT.
SETS HEATING THHRESHOLD FOR LOW
ROOM TEMPERATURE ALARM.
HIGH ROOM TEMPERATURE ALARM
ADJUSTMENT.
SETS HEATING THRESHOLD FOR HIGH
ROOM TEMPERATURE ALARM.