AD21 GB Issue 03
3
MAIN MENU
– QUICK REFERENCE
TEMP
OK
RH%
OK
ALARM HI
OK
ALARM LO
OK
C/F
OK
SAVE
OK
CANCEL
OK
37.5C
OK
-
+
RH 20%
EX ONLY
HI 2.0C
OK
-
+
LOW 3.0C
OK
-
+
DISP C
OK C F
INCUBATION TEMPERATURE.
RANGE 20.0 – 40.0°C (68.0 – 104.0°F).
DEFAULT 37.5°C (99.5°F).
RELATIVE HUMIDITY.
RANGE 20% – 80%.
DEFAULT 20%.
ONLY FOR EX VERSION WITH
ADVANCE HUMIDITY PUMP – SEE
SECTION 8.7.
HIGH TEMPERATURE ALARM.
RANGE 1.0 – 5.0°C (1.8 – 9.0°F) ABOVE
SET INCUBATION TEMPERATURE.
DEFAULT 2.0°C (3.6°F).
LOW TEMPERATURE ALARM.
RANGE 1.0 – 5.0°C (1.8 – 9.0°F) BELOW
SET INCUBATION TEMPERATURE.
DEFAULT 3.0°C (5.4°F).
CELSIUS / FAHRENHEIT DISPLAY.
SWITCHES ALL TEMPERATURE
FIGURES BETWEEN °C AND °F.
DEFAULT °C.
SAVE.
ALL CHANGES ARE SAVED. RETURN
TO NORMAL OPERATION SCREEN.
CANCEL.
ALL CHANGES ARE IGNORED.
RETURN TO NORMAL OPERATION
SCREEN.
PRESS BOTH BUTTONS TO UNLOCK THE MAIN MENU
SELECT THE OPTION / RETURN TO THE MENU.
GO FORWARD ONE SCREEN / INCREASE THE VALUE / DISPLAY IN CELSIUS.
GO BACK ONE SCREEN / DECREASE THE VALUE / DISPLAY IN FAHRENHEIT.
COOLING
OK
MINS OFF
OK
-
+
PERIODIC INCUBATION COOLING.
S
WITCHES HEATER OFF FOR A
TIMED PERIOD ONCE EVERY 24
HOURS (SEE SECTION 11).
RANGE 60 – 360 MINUTES.
DEFAULT - OFF.