BRIGHT 8000 Base Operating, Maintenance And Spare Parts Manual Download Page 7

Issue 3  April 2011  

7

1. INTRODUCTION 

 

 
 

1.1 

Receipt and Unpacking 

 

This instrument received a final test and inspection prior to despatch from the factory. The 
following instructions are given for the re-assembly of the instrument, adjustments and its 
correct use. If the instrument is received before preparations for installation are completed, it 
should be stored in a clean, dry place and not exposed to dirty or damp conditions. 
 
 
 
 

1.1.1 Receipt 

 

Immediately upon receipt of the instrument, make a careful examination for evidence of 
damage encountered in transit. If any damage is found or suspected, notify both the carrier 
and Bright Instrument Company Limited. 
 
 
 
 

1.1.2 Unpacking 

 

All packing must be carefully removed and parts checked against the enclosed packing list. If 
any damage or discrepancy is noted, please inform our agent/distributor, or bright Instrument 
Company Limited direct, immediately. At least two people are required during the unpacking. 
When removing the main instrument from the packaging, grip the instrument on the underside 
(one person to each side) and lift out the microtome. Do not lift by the sledge or attempt to 
separate the sledge from the casting within  the case as damage can result.  
 
 
 
 

1.2 ASSEMBLY 

AND 

INSTALLATION 

 
 
 
 

1.2.1 Positioning 

 

The instrument should be positioned on a level, sturdy surface at a height convenient for the 
operator. It may be used with its long axis parallel to the bench, at right angles or obliquely. 
Ensure it has been positioned away from direct, hot sunlight and is in a location completely 
free from draughts. 
 
 

Summary of Contents for 8000 Base

Page 1: ...Issue 3 April 2011 1 Operating Maintenance and Spare Parts Manual for the Bright 8000 Base Sledge Microtome Serial Number...

Page 2: ...ng 7 1 1 1 Receipt 7 1 1 2 Unpacking 7 1 2 ASSEMBLY AND INSTALLATION 7 1 2 1 Positioning 7 2 OPERATING INSTRUCTIONS 8 2 1 Operating handle 3 8 2 2 Thickness control 5 8 2 3 Coarse advance trim control...

Page 3: ...4 MAINTENANCE 11 4 1 Cleaning 11 4 2 Lubrication 11 4 3 Decontamination 11 4 4 Removing the Sledge 11 MATERIAL SAFETY DATA SHEET 12 Name Microtome Lubricating Oil 12 5 SPECIFICATIONS 13 6 ACCESSORIES...

Page 4: ...ition Remove the knife blade or cover with the guards or visor provided b DO NOT leave unboxed knives blades lying around Place knives blades that are not in use in their boxes or packets c DO NOT car...

Page 5: ...contact with potential pathogenic or radioactive material it is essential that it be decontaminated 2 Set procedures are laid down by the Health and Safety Executive for decontamination For the avoida...

Page 6: ...warranty such warranty shall thereafter be limited to a period of six months from the date when the goods shall have been re delivered to the Buyer c this warranty does not apply to any defects caused...

Page 7: ...packing must be carefully removed and parts checked against the enclosed packing list If any damage or discrepancy is noted please inform our agent distributor or bright Instrument Company Limited dir...

Page 8: ...ed then turn clockwise past the required thickness then anticlockwise 2 3 Coarse advance trim control 4 Quickly raises or lowers the specimen 450 m per turn As this control is always engaged it can be...

Page 9: ...and that the heel of the knife is not caught up on the slot in the knife clamps d Screw the knife clamp levers down just sufficiently to steady the knife in position e Slacken the knife block clamp le...

Page 10: ...built up wax 2 Wax embedded specimens usually cut more easily if they have been chilled It is common practice to trim a batch of wax blocks then place them face down on a cold plate or crushed ice to...

Page 11: ...The Slide Blocks Dovetails Fig 1 6 should also be lubricated on a regular basis with Part No 57425 Microtome Sledge Oil 4 3 Decontamination It is the responsibility of the user to ensure that a decont...

Page 12: ...ake inhalation unlikely at ambient temperatures INGESTION May cause nausea vomiting and diarrhoea Toxic effects can be delayed by up to 48hours OTHER NONE known Respiratory Fumes may irritate respirat...

Page 13: ...mm Maximum specimen size Maximum specimen size 250x110mm Section thickness range 0 040 m in 1 m increments Total feed range 8 horizontal and vertical optional Specimen holder Specimen vice or quick re...

Page 14: ...tage 30x30mm 53024 Solid state freezer a Peltier effect freezing stage 40x40mm 8000 208 Cryostage freezing stage 130x90mm 51719 Object holder 73x54mm 51720 Object holder 100 x 66mm 51836 Object holder...

Reviews: