HS-250 Medical Scale User Instructions
25
7.5.2 Adaptor Voltage
Check the adaptor output cable connecter voltage
l
Voltage must be minimum 9VDC and maximum 15VDC
7.5.3 PCB Input Voltage
Check the PCB input power connector voltage:
l
Voltage must be minimum 9VDC in to the pin AD+
l
Check Battery Voltage and Charging Voltage
1.
Check the Battery Voltage:
Voltage must be minimum 6VDC. If below the 6VDC connect the
adaptor for charging
The battery voltage below the 5.5VDC, replace the battery and install
new 6V/3.4Ah battery.
2.
Check the Battery Charging Voltage:
Remove the battery connection terminals (Red and Black) from the
battery.
Connect the power and turn on the scale
Voltage into the terminal must be a minimum of 6.5VDC