12
Model 405-LCD
The scale weight is tared off between each ingredient to give the operator full viability
of the weigh of each ingredient added.
1.
Zero the scale by pressing the ON/OFF/ZERO key.
2.
Place an empty container on the scale and press the TARE/UNIT key. The
scale’s display will now show a Net Zero reading and the NET annunciator will
now be displayed.
3.
Place the product to be weighed inside the container on the scale.The scale
will display just the weight of the product only and not the weight of the
container.
To remove a Tare, first remove the container from the scale and then press the ON/
OFF/ZERO key. The display will zero and the NET annunciator will turn off. The Zero
Annunciator will then be displayed on.
3.6 Calibration
Before calibration can start, make sure you have the correct calibrated weights to
calibrate the scale and that the scale is on a flat level surface. To calibrate the scale,
you will need the full scale capacity in weights 6kg or 12lb and 15kg or 30lb.
1.
With no load on the scale and in the off position, press and hold down the
TARE/UNIT button and the ON/OFF/ZERO button at the same time until CAL-0
appears on the display. Release the two buttons.
2.
With no weight on the scale, press ON/OFF/ZERO key until “0 in CAL-0” is
flashing.
3.
When CAL-F is displayed, press and hold the TARE/UNIT key to choose the
calibration units of measure (kg or lb). Now load the full capacity of weight onto
the platter and press the ON/OFF/ZERO key. The letter F in CAL-F will start to
flash.
4.
When CAL-0 appears again, remove the weight from the platter and press the
ON/OFF/ZERO key. 0 in CAL-0 will flash.
5.
If the calibration was successful, the scale will automatically return to normal
weighing mode. If CAL-Er is displayed, there is an error in calibration and the
scale needs to be re-calibrated.
6.
To exit calibration mode, press and hold the ON/OFF/ZERO key.
3.7 AUTO-OFF and BACKLIGHT settings
Auto-Off Setting
If the scale has no movement or key presses in this period of time, the scale will
automatically turn off to save battery life.
NOTE: TARE range for this scale is up to 100% full scale capacity.