Doc. No. 003259 En Issue: X-1
BREAS MEDICAL
2–7
PV 101+/PV 102+ Service Manual
MAINTENANCE SERVICE INSTRUCTIONS
2.6
Final inspection before delivery
2.6.1
Function check
1
Connect the power cord to the power socket of the PV 101+/PV 102+.
2
Connect the patient circuit.
3
Start the PV 101+/PV 102+ and check that it functions normally.
4
Disconnect the power cord and check that the power failure alarm sounds. Recon-
nect the power cord. The alarm should be silenced.
5
Switch on the PV 101+/PV 102+. Upon starting, all the LEDs on the front panel
should be lit and the trigger settings should be indicated on the bar graph. The
entire bar graph should then become yellow for 2 seconds before the device starts
functioning normally.
6
(PV 102+ only) Switch on the PV 102+ and wait for 15 seconds for the leakage
alarm to be activated.
7
Switch off the device.
2.6.2
Checking the low pressure leakage alarm (PV 102+)
1
Adjust the pressure settings to:
IPAP 10
mbar
EPAP 10
mbar
2
Make sure that the alarm LED is lit and that an audible alarm is heard.
2.6.3
Checking the trigger function
1
Adjust the pressure settings to:
IPAP 10
mbar
EPAP 4
mbar
2
Connect the patient tube without a mask.
3
Block the air outlet on the tube. When you remove the blockage, a triggered
breath should be initiated.
2.6.4
Checking the pressure and rate
Pressure
The measurement should be made using a blocked patient circuit with leakage provided
(see figure below).
1
Adjust the pressure settings to:
IPAP 10
mbar
EPAP 10
mbar
2
Measure the pressure and check that it is correct (tolerance: ±4% of the maximum
value and ±8% of the current pressure value).
Summary of Contents for PV 101+
Page 76: ...FAULT TRACING PV 101 PV 102 Service Manual 8 2 BREAS MEDICAL Doc No 003259 En Issue X 1...
Page 78: ...9 2 BREAS MEDICAL Doc No 003259 En Issue X 1 APPENDICES PV 101 PV 102 Service Manual...
Page 82: ...9 6 BREAS MEDICAL Doc No 003259 En Issue X 1 APPENDICES PV 101 PV 102 Service Manual...
Page 87: ...BREAS MEDICAL 10 PV 101 OPERATING MANUAL...
Page 88: ...BREAS MEDICAL...
Page 89: ...BREAS MEDICAL 11 PV 102 OPERATING MANUAL...
Page 90: ...BREAS MEDICAL...