Charging the Battery
1.
With both the
Z2
and the
Bat-
tery
fully inserted into the
Power-
Shell
, plug the
Power Supply
into
the
PowerShell
as shown and into
a power outlet.
2.
Ensure the
Z2
is in standby mode
(display reads 0.0). The
Battery
will not charge unless the
Z2
is in
standby mode. The
Battery
status
indicator on the
Z2
will flash when
the
Battery
is charging. The number
of solid bars indicates the current
level of charge. The battery status
indicator will stop flashing when the
Battery
is fully charged.
WARNING
: Only charge the
Battery
with the supplied
Breas Medical
PowerShell
and the supplied Breas Medical
Power Supply.
Do not use any other battery charger, AC
adapter, or other device to charge the
Battery
.
CAUTION:
Make sure your
Battery
is fully charged
before commencing CPAP therapy using the
PowerShell
.
The battery life of a fully charged
Extended Life Battery
in
the
PowerShell
is typically 8.9 hours or longer, after which
treatment will terminate.