PowerShell

®

 with  

Extended Life Battery 

Getting Started Guide

Summary of Contents for PowerShell

Page 1: ...PowerShell with Extended Life Battery Getting Started Guide ...

Page 2: ...y with a hammer step on the Battery or otherwise subject it to strong impacts or shocks Do not solder onto the Battery Do not expose the Battery to water or salt water or allow the Battery to get wet Do not disassemble or modify the Battery The Battery contains safety and protection devices which if damaged may cause the Battery to generate heat rupture or ignite Do not place the Battery on or nea...

Page 3: ... attach the Battery directly to a vehicle cigarette lighter or to the Power Supply plug Do not continue charging the Battery if it does not recharge within the specified charging time Doing so may cause the Battery to become hot rupture or ignite Remove the Battery from the PowerShell if the Z2 is not likely to be used for some time Do not use the Battery to power any device other than the Z2 Only...

Page 4: ...ions on using the Z2 CPAP The PowerShell is also fully compatible with the Z1 CPAP system PowerShell with Neoprene Sleeve Extended Life Battery with Shuttle Package Contents WARNING The Z2 must only be used with the optional Breas Medical PowerShell with Breas Medical Battery or the Breas Medical Power Supply DO NOT use any other power source with the Z2 PowerShell Getting Started Guide ...

Page 5: ... that it has locked into place The Z2 will power on in standby mode The display will read CAUTION Insert the Z2 carefully ensuring the power barrel connector slides into the power input on the Z2 Instructions for Use NOTE The Z2 can be removed from the PowerShell by pressing the two symbols on the sides of the PowerShell and sliding the Z2 out NOTE Remove the Battery from the PowerShell when the Z...

Page 6: ...order of pieces are as follows PowerShell tube adapter 8 tube Qtube 4 tube HME mask 4 Push the start stop button to initiate therapy Push this button a second time to stop therapy Alternatively the Z2 Auto fea tures an Auto On Off which starts therapy when you place the mask on your face and start to breathe and stops therapy when the mask is removed NOTE It is recommended that you fully charge th...

Page 7: ...lid bars indicates the current level of charge The battery status indicator will stop flashing when the Battery is fully charged WARNING Only charge the Battery with the supplied Breas Medical PowerShell and the supplied Breas Medical Power Supply Do not use any other battery charger AC adapter or other device to charge the Battery CAUTION Make sure your Battery is fully charged before commencing ...

Page 8: ...2 4 Make sure to fully charge the Battery for 12 hours after use 5 The battery icon on the Z2 screen will stop flashing and all bars will be solid NOTE Battery icon starts showing 4 bars at 90 charge Wait for battery icon on Z2 LCD screen to stop flashing Number of Solid Bars Charge Remaining 1 10 35 2 35 63 3 63 90 4 90 100 Typically it will take 12 hours to fully charge the battery After a few m...

Page 9: ...p on the Z2 To Release Overnight Battery from PowerShell Press and hold the 2 blue buttons on the white cap of the battery and with even pressure pull up at the same time Using the DC Mobile Adapter Your PowerShell Battery can be charged with the DC Mobile Adapt er using most automobile charging systems The amount of time will vary depending on the voltage delivered by the car battery Some cars ch...

Page 10: ...ltitude and pressure set ting of the unit Higher altitudes and higher pressures will use more battery life Temperature can also be a factor Colder weather will use more battery life The battery is designed to be used in temperatures between 41 F and 95 F Do not charge the battery in temperatures above 80 F ...

Page 11: ...ating times Capacity 99 4 Watt hours Lifetime approximately 300 charge discharge cycles Shelf life Batteries are shipped with between 30 and 50 of rated capacity providing a minimum of 6 months shelf life when stored at 25 C If storage temperature exceeds 25 C over a 6 month period then shelf life will be reduced and the battery should be recharged periodically Storage and transport temperature 4º...

Page 12: ...es of America Breas Medical Inc N Billerica MA 01862 USA PowerShell and Z1 are registered trademarks of Breas Medical Inc Z2 is trademark of Breas Medical Inc Breas and the Breas logo are trademarks of Breas AB US Patent Pending Copyright 2014 2019 Breas Medical Inc All rights reserved ...