8
9
English
Care and cleaning
The probe tip is the most delicate part of the thermometer.
It must be clean and intact to ensure accurate readings.
If the thermometer is ever accidentally used without a lens filter,
clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with
alcohol. After the alcohol has completely dried out, you can put a new lens
filter on and take a temperature measurement.
If the probe tip is damaged, please contact your local service centre.
Use a soft, dry cloth to clean the thermometer display and exterior. Do not use
abrasive cleaners. Never submerge this thermometer in water or any other liquid.
Store thermometer and lens filters in a dry location free from dust and contamination
and away from direct sunlight.
Additional lens filters (LF40) are available at most stores carrying Braun ThermoScan
®
.
Insert new batteries when the battery symbol appears
on the display.
Open the battery compartment. Remove the batteries and replace
with new batteries, making sure the poles are in the right direction.
Snap battery door into place.
1
2
3
Only discard empty batteries. To protect the environment, dispose of empty
batteries at your retail store or at appropriate collection sites according to
national or local regulations.
Calibration
The thermometer is initially calibrated at the time of manufacture. If this thermometer
is used according to the use instructions, periodic re-adjustment is not required. If at
any time you question the accuracy of the temperature measurement, please contact
your authorized service center.
Manufacturing date is given by the LOT number located in the battery compartment.
The first three (3) digits after LOT represent the Julian date that the product was
manufactured and the next two (2) digits represent the last two numbers of the
calendar year the product was manufactured. The last identifiers are the letters that
represent the manufacturer.
An example: LOT 11614K, this product has been manufactured on the 116th day of the
year 2014.
Replacing the batteries
Errors and troubleshooting
Situation
No lens filter is attached.
The thermometer probe was not
positioned securely in the ear.
An accurate measurement
was not possible.
POS = position error
Ambient temperature is not within
the allowed operating range
(10 – 40 °C or 50 – 104 °F)
.
Temperature taken is not within
typical human temperature range
(34 – 42.2 °C or 93.2 – 108 °F)
.
HI = too high
LO = too low
System error – self-check display flashes
continuously and is not followed by the
ready beep and the ready symbol.
If error persists,
If error still persists,
Battery is low, but thermometer will still
operate correctly.
Battery is too low to take correct
temperature measurement.
Error Message
Solution
Attach new, clean lens filter.
Make sure that the positioning of the
probe is correct and remains stable.
Replace lens filter and reposition.
Press Start button to begin a new
measurement.
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between
10
and 40 °C or 50 and 104 °F
.
Make sure the probe tip and lens are
clean and a new, clean lens filter is
attached. Make sure the thermometer
is properly inserted. Then, take a new
temperature.
Wait 1 minute until the thermometer
turns off automatically, then turn
on again.
... reset the thermometer by removing
the batteries and putting them back in.
… please contact your local service
centre.
Insert new batteries.
Insert new batteries.