This product conforms to the provisions of the EC directive 93/42/EEC (Medical Device
Directive). This device conforms to the following standards:
•
EN 60601-1: 2006 + AC:2010: - General requirements for basic safety and essential
performance
•
EN 60601-1-2:2007 - Electromagnetic compatibility requirements and tests
•
EN 60601-1-11:2010 - Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment
•
EN 1060-1:1995 + A2:2009 – Non-invasive sphygmomanometers - general
requirements
•
EN 1060-3:1997 + A2:2009 - Non-invasive sphygmomanometers - Supplementary
requirements for electro-mechanical blood pressure measuring systems.
•
EN 1060-4:2004 - Non-invasive sphygmomanometers - Test procedures to determine
the overall system accuracy of automated non-invasive sphygmomanometers.
•
IEC60601-1-11:2010-General requirements for basic safety and essential performance-
Collateral Standard: Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment
Medical electrical equipment needs special precautions regarding EMC.
For detailed description of EMC requirements please contact an authorized local service centre
Portable and mobile RF communications equipment can affect medical electrical equipment.
Please do not dispose of the product in the household waste at the end of its useful life.
Disposal can take place at your local retailer or at appropriate collection points provided
in your country.
Warranty
This product comes with limited warranty commencing on the date of purchase. Within the warranty
period we will eliminate, free of charge, any defects in the appliance resulting from faults in materials
or workmanship, by replacing the complete appliance.
This warranty is applicable only for the appliance supplied by the appointed distributor.
This warranty does not cover: damage due to improper use, normal wear or use as well as defects
that have a negligible effect on the value or operation of the appliance. The warranty becomes void
if repairs are undertaken by unauthorized persons and if original Braun parts are not used. To obtain
service within the warranty period, please contact us.
HK Listing No.: 130467
Specifications
Method of measurement
Oscillometric
Model number
BP 6200
Range of measurement
Pressure 0~300 mmHg
Pulse 40~199 beats/minute
Accuracy
Pr/- 3 mmHg
Pulse +/- 5 % Max.
Inflation
Deluxe automatic
Display
Liquid crystal display – systolic, diastolic, pulse rate
Backlight Display
Sets of memory
60 sets per user
Cuff size
Small cuff = 22-32cm arm circumference
Large cuff = 32-42cm arm circumference
Operating temperature
+10 °C ~ + 40 °C, less than 85 % R.H.
Storage temperature
-20 °C ~ +60 °C, less than 85 % R.H.
Operating atmospheric pressure
860-1060 hPa
Unit weight
Approximately 500gr (without batteries)
Power supply
Alkaline battery: 4 x AA (LR6) 1.5V
Battery life
300 times measurement
Auto power off
Whenever not used for 1 minute
Accessories
4 batteries, 2 arm cuffs with tube, instruction manual, pouch,
travel bag
Service life
5 years
IMPORTANT
Read the operating instructions.
If device is not used within specified temperature, humidity and atmospheric pressure
ranges the technical accuracy of the measurement cannot not be guaranteed.
Classification:
•
Internally powered equipment
•
Type BF equipment
•
IP22: Protected against solid foreign objects of 12.5 mm diameter and greater.
Protected against vertically falling water drops when the device is tilted up to 15°
•
Not suitable for use in the presence of flammable anesthetic mixture with air, oxygen
or nitrous oxide
•
Continuous operation with short-time loading
Operating temperature
Storage temperature
Storage humidity
0297