Page 5
Lift Terminology
Operation Notes section (pages
17 and 18).
Automatic Inboard Roll Stop
and Bridge Plate: CL917 lift
models are equipped with an
automatic inboard roll stop that
also serves as the bridge plate.
The bridge plate bridges the gap
between the lift platform and the
vehicle floor. The roll stop/bridge
plate automatically rotates from
the horizontal (bridging) position
to the vertical (roll stop) position
as the lift lowers and raises.
Further details regarding the
automatic mechanical inboard
roll stop are provided on pages
16 and 17.
Pump Module
Indicator Light
Power/Interlock Indicator Light:
A green indicator light is built
into the pump module that
illuminates when remote
interlock(s) are made (circuit
complete), there is power to
the lift and the interlocking
handrail belt is buckled. All
three conditions must be met.
The light will shut off if the
handrail belt is unbuckled (lift
functions are disabled also).
An illuminated light indicates
the lift is operational.