background image

∙ Sync Mode: 

When Sync is set to ON, Apical Action will be    

  activated.

∙ Apical Action: 

Apical Reverse or Apical Stop can be set.   

  The file will either stop or reverse when the file tip reaches    
  the apex. 

∙ Auto Start/Stop: 

When this is set to ON, the file starts  

  rotation when it is inserted and stops when it is removed  
  from the canal.  

∙ Speed: 

Use 

+ or – Buttons

 to adjust to desired  speed

***

∙ Reverse Torque:

 Use 

+ or – Buttons

 to adjust to desired  

  torque

***

Main Switch

Memory Setting

Cutting Direction

Battery Power

Rotation Mode

Speed

Torque

Select Key

(—) Button (+) Button

Torque Display

Current 

Torque

Reversing

Torque

BASIC OPERATION

∙ Turn ON: 

Press the 

Main Switch

∙ Start/Stop the Motor:

 Press the 

Main Switch 

again.

∙ Select/Change Memory Setting:

 Press the 

+ or – Buttons

 to cycle between the memory setting (M1-M6)

∙ Turn OFF:

 Press the 

Main Switch

 and the 

Select Key

 at the same time

PREFERENCE MENU

1.

  With the EndoSync Plus turned off, hold the 

Select Key

 and press the 

Main Switch

 to access the preference menu.

2.

  Tap the 

Select Key

 to cycle through the different options to customize. Use 

+ or – Buttons

 to change the settings. 

 

∙ Beep Volume:

 High, Low, OFF

 

∙ Auto Power Off:

 1 min-15 min (idle time before the device powers off)

 

∙ Display Type:

 Positive or Negative (inverted color scheme)

 

∙ Dominate Hand:

 Right or Left

 

∙ Back Light color Change:

 Adjusts background color from colored to black + white

 

∙ Stand by Screen Return (SSR):

 1 sec- 15 sec. Set the time that elapses before the display reverts back  

 

  to the standby screen 

 

∙ Setting Lock: 

ON or OFF, locks the settings in place for M1-M6 (speed, torque, direction etc.)

MEMORY SETTINGS OPERATION

∙ Rotation Mode:

 OTR, CW (FWD), CCW

*

 (REV)

 

 

    a.  If OTR is selected, choose OTR Cut  

 

 

 

         Direction: CW or CCW

 

 

    b.  If OTR, select desired Cut Angle and    

 

                     Non-cutting angle

 

 

    c.  Cutting Angle: 90º

**

, 150º, 180º,

 

 

         270º, 360º

 

 

    d.  Non Cutting Angle: 30º, 60º 90º, 12

0

º

Hold down the 

Select Key

 for 2 seconds to change the memory settings. Tap the 

Select Key

 to cycle through the 

different memory setting options.

ADVANCED SETTINGS FOR USE WITH THE ENDOSYNC A.I. APEX LOCATOR

Hold down the 

Select Key

 for 2 seconds to change the memory settings. Tap the 

Select Key

 to cycle through the basic settings 

until you see “Sync”.  Use 

+ or – Buttons

 change the settings for each option below.

*should only be used with files that cut on the Reverse or CCW direction
**90 degree cut angle only achievable at 100rpms
***Some speed and torque settings not available in OTR mode

Cannot be used in combination with Apical Torque Reduction

††

Cannot be used in combination with Apical Slow down or Torque Slow Down

†††

Cannot be used in combination with Apical Torque reduction

Please reference page 15 of the operation manual for more details.

Below settings are only available when OTR is disabled:

∙ Apical Slow Down

When ON, the file slows down as it  

  approaches the apex. 

∙ Apical Torque Reduction

††

When ON, the torque setting  

  that triggers reverse rotation is reduced as the file tip  
  approaches the apex. apex. 

∙ Torque Slow Down

†††

When ON, the motor speed will  

  slow down as the torque load increases.

Reviews: