10
1. Insert battery pack into a charging bay receptacle.
2. Follow warnings, cautions and notes in BSPMAX™ II
PM-X00-710 (9.6V) & PM-X00-715 (12V) Battery
Packs Instructions for Use.
3. A single light will illuminate to indicate battery pack
status.
• All pulse - System is powered on.
• Yellow (pulsing) - Primary diagnostic test to determine voltage level.
• Yellow (solid) - Battery Pack is charging.
• Green (solid) - Battery Pack is charged and ready for use.
• Red (pulsing) - Secondary diagnostic test to determine charging capacity.
• Red (solid) - Charging cycle failed and battery pack should be disposed of
properly (refer to Battery IFU for Product Disposal Information).
Battery Pack Charging
Charging
Bay
Receptacle
Battery
Pack
CHARGE (Yellow)
READY (Green)
SERVICE (Red)