20
The BSPMAX™ II Battery Charger is intended for use in the electromagnetic environment specified below.
The user of the BSPMAX™ II Battery Charger should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 KV contact
±8 KV air
±2, 4, 6 KV contact
±2, 4, 8 KV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 KV for power
supply lines
±1 KV for
input/output lines
±2 KV for power
supply lines
±1 KV for
input/output lines
Main power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 KV
differential mode
±2 KV
common mode
±1 KV
differential mode
±2 KV
common mode
Main power quality should be that of a typical
commercial or hospital environment.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
Specifications (continued)
5. Electromagnetic Compatibility Requirements continued