4-22
100-214-276 Rev. 3
Chapter 4: Installation and Setup
Installation Steps
3.
Lift the actuator assembly into position on your mount, and secure using the metric bolts provided.
CAUTION
!
In the event you must use bolts of a different length, ensure that the bolts extend more than
0.25 inch (6 mm) into the threads in the actuator housing, but less than 0.40 inch (10 mm).
4.5.4 Mount the Power Supply
The power supply is designed to be placed on a workbench (rubber feet on bottom) within cable-length lim-
its of the actuator, or it may be rack-mounted in a standard 19-inch Rack (using an optional rack mount han-
dle kit). It has two rear-mounted fans which draw cooling air from rear to front, which must be free from
obstruction. Do not place the power supply on the floor or in other locations that will allow dust, dirt or con-
taminants to be drawn into the power supply.
The controls on the front of the power supply must be accessible and readable for setup changes.
All electrical connections are made to the rear of the power supply, which should be positioned in your
workspace with adequate clearance (approximately 4 inches or more on either side, and 6 inches to the rear)
for cable access and ventilation. Do not place anything on top of the power supply case.
In the event the system is to be installed in a high dust environment, the use of a fan filter kit (101-063-614)
is required.
See
for a dimensional drawing of the 2000X Power Supply.
Figure 4.15
Connections on Rear of Power Supply
The cable lengths are limited based on the operating frequency of the welding system. Performance and
results can suffer if the RF cable is crushed, pinched, damaged or modified. Contact your Branson Repre-
sentative if you have special cable requirements. In some cases, remote operation from a User I/O or a Host
Computer can be used to solve a distance limitation.
Fans
RF Connector
(ultrasonic energy out)
Power Cord
Parallel Printer Connector
Actuator Interface Connector
Model Data Tag
User I/O Connector
Serial Port (RS-232) Connector for Host Computer
USER I/O
J3
SW1
ACTU ATOR
SIGNALS
J7
HOST
TERM
J9
J8
MODEL
PRINTER
THIS EQUIPMENT COMPLIES WITH
F.C.C. REGUL
ATIONS PART 18
EDP NO.
WARNING
THIS EQUIPMENT COMPLIES WITH
F.C.C. REGUL ATIONS PART 18
GROUP 2 CLASS A ISM
ULTRASONICS CORPOR ATION
41 EAGLE ROAD, DANBU RY, CT 06813-1961
MAX POWER
Serial No.
Year & Month of Manufacture
INPUT
DIP Switch for User I/O
Ethernet Connector (Optional)
DOC EXPIRES 12PM 7/24/2012. Article or Material must comply with the requirements
stipulated by RoHS in its current version