5.3
Viewing the Current Status of the PTU system
From the “MENU” button, select “General” from the status menu. This shows
information such as the current output time, UTC time, up time, the current input
reference, the current output format, current system state, alarm status, battery
status, and current internal temperature.
Figure 4 - PTU Status Page
Output Time
: The current time of day being displayed or being output by the PTU
UTC Time
: The current time of day from the GPS constellation in Universal Time
Coordinate (UTC)
Up Time
: The amount of time that the PTU has been powered on
Hour Meter
: This is the number of hours since the PTU was first built.
Reference
: The current reference input being used by the PTU
Output TimeCode
: The current timecode format in use by the PTU’s outputs
TFOM
: Time Figure Of Merit – the difference between the unit’s time and the
reference input. A 1 means the unit is accurate to within 1ns, a 2 means the unit is
accurate to within 10ns, a 3 means 100ns, etc
System state
: The current mode of the PTU’s internal clock