System User Guide Rev. 1.6 VectorVision
97
CLEANING
8.3
VectorVision Components
8.3.1
VectorVision² and VectorVision compact Bases
Illustration
Figure 48)
How to Clean
Risk of electrical shock: prior to and during all cleaning and disinfection work, unplug
power supply and prevent system from being switched on again.
Do Not Use
• Acid solvents
• Caustic solvents with pH value higher than 9.5
To prevent damage to the device, only use approved solutions.
No.
Component
a
VectorVision² base
s
VectorVision compact base
a
s
Steps
1.
Ensure that system is in parking position.
2.
• Clean surfaces using a surface disinfectant such as Meliseptol.
• Follow disinfectant manufacturer’s recommendations.