Port
Example
Handling
USB
USB ports for connecting USB 2.0 and 3.0 devices for
patient data transfer.
NOTE: Do not connect network adapters or USB ex-
tension cables to the USB ports.
Hospital net-
work
Galvanic isolated Ethernet port for connecting to Hospi-
tal network.
There is a latch at the bottom which locks when plug-
ged into the port. When removing the connector, press
the latch, then pull plug out.
Feature Panel
Microscope
Only connect microscopes via a Brainlab microscope
cable. Plug in the cable by connecting all connectors
(DVI-I, USB, CVBS) to the corresponding port.
For detailed information on Microscope Integration see
the corresponding
Instrument User Guide
.
DisplayPort
Connect DisplayPort display using DisplayPort 1.2 ca-
ble (max. 5m for 1920 x 1080/60 Hz and max 3.6 m for
3840 x 2160/30 Hz/60 Hz).
Related Links
6.1.1 Digital OR Computer For Barco Integration Connections on page 61