Fluoro Registration Kit for Ziehm Vision FD Vario 3D 

55730 

       

 

  

 

 

 

 

 

1a 1b 

ENGLISH 

 

This Fluoro Registration Kit can be used to register 3D images 
acquired with a Ziehm Vision FD Vario 3D C-arm. 

1. Preparing the Fluorostars 

Tightly screw the reflective marker spheres by hand onto each 
attachment pin of the Fluorostar. 

Determine which two of the three interfaces on the C-arm head 
are best oriented towards the camera in the 3D scan start 
position, for attachment of the Fluorostars. 

2. Attaching the Fluorostars 

Gently push back the safety clip and open the lever of the 

Fluorostar, attach the star onto the interface of the C-arm head 
and close the lever. Repeat the process for the second 
Fluorostar. 

3. Draping: 

Place the drape over the Fluoro Registration Kit and the C-arm 

head according to the instructions provided by the C-arm and 
drape manufacturers. To ensure that the Fluoro Registration Kit 
is fully visible to the camera, prevent the drape from wrinkling 

by securing it to the Fluorostar using a sterile rubber band on 
the rear of the Fluorostar. 

 
Warning: When draping the C-arm and Fluoro Registration 
Kit, the drape should be as smooth as possible. The 
reflective marker spheres should not be covered by 
wrinkles or non-transparent parts of the drape, such as 
seams or stickers. 
 
Warning: Do not attempt to mount the Fluorostar on top of 
the drape. It must be mounted before draping. Do not 
sterilize the Fluorostars. Use only the approved drape: 
Ziehm Art. No. 13034 / Heisig Art. No. 34132 with sterile 
rubber bands (Ziehm Art. No. 131160). The 12” drape 
should only be used if all three reference geometries are 
mounted on the C-arm (Ziehm Art. No. 13031 / Heisig Art. 
No. 34172 with sterile rubber bands (Ziehm Art. No. 
131160)). 
 
Warning: Plan the setup of the OR prior to surgery. The 
camera must have an unobstructed view of the reflective 
marker spheres on the Fluorostars and instruments to be 
able to navigate. Markers must be completely visible to the 
camera and must not be damaged, masked, moistened or 
contaminated, e.g. with blood. 
 
Warning: The reflective marker spheres must be 
exchanged if the surface becomes damaged or 
contaminated. Only use new disposable reflective marker 
spheres, and do not re-use them. 
 
Warning: The Fluoro Registration Kit is a highly accurate 
and sensitive medical device. Handle it with extreme care. 
If you drop or otherwise damage it, return it immediately to 
BrainLAB for testing, replacement or repair. Failure to do 
so may lead to serious patient injury. 
 
Warning: Do not rotate the C-arm with the mounted Fluoro 
Registration Kit under the OR-table. 
 
Warning: Always apply the brakes when the Fluorostars are 
attached. If the brakes cannot be applied, secure the C-
arm manually. 

 

Warning: The Fluorostars must be removed from the C-arm 
after use. Place Fluorostars in the corresponding storage 
case to avoid damage and malfunction. 
 
Warning: Always check that the Fluorostars have been 
attached properly and that the connection is rigid and 
precisely fitted.  
 
Warning: The device shall be clean before use. The 
attachment pins of the shifting markers must be thoroughly 
cleaned to avoid incorrect navigation and patient injury. 
 
For further information please see BrainLAB user manual 
or consult BrainLAB customer support. 

 