bpr Dialflow Regulator Instructions For Use Manual Download Page 7

 

 

Applied Standards 

AS 2473.3 

Valves for compressed gas cylinders 

Part 3: Outlet connections for medical gases (including pin-indexed yoke 
connections) 

AS 3840.1 

Pressure regulators for use with medical gases 

Part 1: Pressure regulators and pressure regulators with flow-metering 
devices 

BS EN ISO 10524-1 

Pressure regulators for use with medical gases. Pressure regulators and 

pressure regulators with flow-metering devices 

BS EN ISO 14971 

Medical devices. Application of risk management to medical devices 

BS EN ISO 15001 

Anaesthetic and respiratory equipment. Compatibility with oxygen 

BS EN ISO 15002 

Flow-metering devices for connection to terminal units of medical gas 
pipeline systems 

BS EN ISO 15223-1 

Medical devices. Symbols to be used with medical devices labels, 
labelling and information to be supplied. General requirements 

BS EN 13544-2+A1 

Respiratory therapy equipment. Tubing and connectors 

BS 5682 

Specification for probes (quick connectors) for use with medical gas 
pipeline systems 

 7.

 

Troubleshooting 

Fault 

Possible Cause 

Solution 

No gas flow 

Gas cylinder turned off 

Check gas supply 

Dial-Flow Regulator not 
connected properly 

Check gas specific probe is correctly 
connected 

Gas cylinder empty 

Replace gas cylinder 

Filter blocked 

Service or repair is required 

Flow selection dial set to ‘0’ 

Select a positive flow rate 

Internal/external leak 

Seal failure 

Service or repair is required 

Insufficient gas flow 

Filter partially occluded 

Service is required 

Supply pressure too low and/or 
gas cylinder nearly empty 

Check gas supply and/or replace gas 
cylinder 

 8.

 

Parts and Spares List 

Part Number 

Description 

818-0040 

Dial-Flow Regulator - O

Range A - Pin Index/Barb 

818-0041 

Dial-Flow Regulator - O

2

 Range C - Pin Index/Barb 

818-0043 

Dial-Flow Regulator - O

2

 Range E - Pin Index/Barb 

Summary of Contents for Dialflow Regulator

Page 1: ...702 0084 14 October 2014 Dialflow Regulator Instructions for Use ...

Page 2: ...ug and may be used in critical treatments Medical gases should only be used for medical purposes on the authority of a physician or authorised clinician and then strictly in accordance with their instructions Oxygen is not flammable however an oxygen enriched atmosphere will drastically increase the rate and severity of combustion Oil and or grease in the presence of an oxygen enriched atmosphere ...

Page 3: ...all stand or cart Do not use or store oxygen equipment near excessive heat 50 C or 125 F or open flame Only appropriately trained personnel working in controlled conditions must perform disassembly assembly and testing of regulators and flowmeters Dialflow Regulators are not MRI compatible Flowmeters may be affected if the cylinder pressure is other than stated in the specifications The threaded b...

Page 4: ...nd these may be different from one country to another ensure that this regulator is the correct connector for the cylinder to be used Once connected to an oxygen cylinder and with the cylinder valve opened the user may select the flow rate prescribed by rotating the gas flow selection dial until the correct flow rate is seen in the viewing window The rate at which gas flows from the regulator is c...

Page 5: ...4 4 After Use Upon completion of the therapy turn the flow selection dial to zero close the cylinder valve and disconnect the outlet hose from the regulator and the patient To remove the regulator from the cylinder first ensure the cylinder valve is closed Bleed off any residual gas pressure in the regulator by turning the flow selection dial to any flow setting until flow ceases Remove the regula...

Page 6: ... service in accordance with the instructions given in the Service Manual or return the device to a recognised BPR Medical Service Centre on or before the date shown 6 Specification Specification Value Flow Accuracy 15 of setting for flow rates above 1 L min 20 10 of settings of 1 L min and below Input Pressure Range 900 to 20 000 kPa 9 bar to 200 bar Flow Settings l min Range A 0 0 02 0 03 0 05 0 ...

Page 7: ...ation to be supplied General requirements BS EN 13544 2 A1 Respiratory therapy equipment Tubing and connectors BS 5682 Specification for probes quick connectors for use with medical gas pipeline systems 7 Troubleshooting Fault Possible Cause Solution No gas flow Gas cylinder turned off Check gas supply Dial Flow Regulator not connected properly Check gas specific probe is correctly connected Gas c...

Page 8: ...cing 212 0012 Pin Index Yoke Seal 212 0048 Bullnose seal Benelux 303 0030 Cylinder Gauge Boot 303 0060 Swivel Cylinder Gauge 0 3mm Restriction 9 Distributor Details Australia BOC Healthcare Customer Service Centre 10 Julius Avenue North Ryde Sydney NSW 2113 T 61 0 2 8874 4400 F 61 0 2 9886 9000 www boc com au New Zealand BOC Healthcare Customer Service Centre 970 992 Great South Road Penrose 1061 ...