bpr Dialflow Regulator Instructions For Use Manual Download Page 3

 

 

 

The use of the BPR Medical oxygen regulators for gases other than compressed oxygen 
is expressly prohibited. 

 

Never permit compressed medical gases to enter a regulator suddenly.  Always open the 
cylinder valve slowly. 

 

Never use medical gases from a cylinder without reducing the pressure through a suitable 
regulator intended for that gas. 

 

Always close the cylinder valve when the regulator is not in use. 

  Never use oxygen as a pressure medium to purge obstructed pipelines or equipment, 

operate pneumatic tools or to build up any pressure in tanks. 

  Never install a pin index regulator Dialflow Regulator with more than one yoke seal 

between the valve and the regulator.  Before attaching the regulator, verify that the post 
valve is not already fitted with a yoke seal. 

 

Ensure that you have the correct regulator for the type of cylinder you are intending to 
use.  Never attempt to fit a regulator to an incompatible cylinder. 

 

Ensure that the threaded fittings or indexing pins on the regulator are properly mated and 
tightened before opening the cylinder valve. 

 

Do not stand in front of a regulator outlet when opening the cylinder valve. 

 

Before removing a regulator from a cylinder, fully close the cylinder valve and release all 
gas from the regulator. 

 

Secure cylinders to a wall, stand or cart. 

 

Do not use or store oxygen equipment near excessive heat (>50 °C or 125 °F) or open 
flame. 

  Only appropriately trained personnel working in controlled conditions must perform 

disassembly, assembly and testing of regulators and flowmeters. 

 

Dialflow Regulators are not MRI compatible. 

  Flowmeters may be affected if the cylinder pressure is other than stated in the 

specifications. 

 

The threaded bull nose cylinder connection for medical oxygen and medical air may be 
the same, ensure that you have the correct cylinder for the application.  This regulator is 
only intended for use with oxygen. 

 

The holes in the side of the body of the device are for venting gas in the event that the 
relief valve is activated.  Do not obstruct these holes or interfere with the relief valve in 
any way. 

  These regulators are designed for use for cylinder pressures up to 20 000 kPa 

(3000 psi/200 bar).  Do not attempt to connect BPR regulators to cylinders having fill 
pressures in excess of this value. 

 

Do not submerge the device in any fluid.  Ensure that no fluid is allowed to enter the inlet 
valve or the vent holes. 

 

Ensure the device is not set between flow settings.  If set between flow settings failure to 
provide therapy may occur.  Display of flow value does not indicate flow is occurring. 

2.2. Cautions! 

 

Device performance may be affected if it is stored or transported in temperature outside of 
the range -20 °C to +60 °C (-4 °F to +140 °F). 

 

The Dialflow Regulator is not suitable for autoclaving. 

Summary of Contents for Dialflow Regulator

Page 1: ...702 0084 14 October 2014 Dialflow Regulator Instructions for Use ...

Page 2: ...ug and may be used in critical treatments Medical gases should only be used for medical purposes on the authority of a physician or authorised clinician and then strictly in accordance with their instructions Oxygen is not flammable however an oxygen enriched atmosphere will drastically increase the rate and severity of combustion Oil and or grease in the presence of an oxygen enriched atmosphere ...

Page 3: ...all stand or cart Do not use or store oxygen equipment near excessive heat 50 C or 125 F or open flame Only appropriately trained personnel working in controlled conditions must perform disassembly assembly and testing of regulators and flowmeters Dialflow Regulators are not MRI compatible Flowmeters may be affected if the cylinder pressure is other than stated in the specifications The threaded b...

Page 4: ...nd these may be different from one country to another ensure that this regulator is the correct connector for the cylinder to be used Once connected to an oxygen cylinder and with the cylinder valve opened the user may select the flow rate prescribed by rotating the gas flow selection dial until the correct flow rate is seen in the viewing window The rate at which gas flows from the regulator is c...

Page 5: ...4 4 After Use Upon completion of the therapy turn the flow selection dial to zero close the cylinder valve and disconnect the outlet hose from the regulator and the patient To remove the regulator from the cylinder first ensure the cylinder valve is closed Bleed off any residual gas pressure in the regulator by turning the flow selection dial to any flow setting until flow ceases Remove the regula...

Page 6: ... service in accordance with the instructions given in the Service Manual or return the device to a recognised BPR Medical Service Centre on or before the date shown 6 Specification Specification Value Flow Accuracy 15 of setting for flow rates above 1 L min 20 10 of settings of 1 L min and below Input Pressure Range 900 to 20 000 kPa 9 bar to 200 bar Flow Settings l min Range A 0 0 02 0 03 0 05 0 ...

Page 7: ...ation to be supplied General requirements BS EN 13544 2 A1 Respiratory therapy equipment Tubing and connectors BS 5682 Specification for probes quick connectors for use with medical gas pipeline systems 7 Troubleshooting Fault Possible Cause Solution No gas flow Gas cylinder turned off Check gas supply Dial Flow Regulator not connected properly Check gas specific probe is correctly connected Gas c...

Page 8: ...cing 212 0012 Pin Index Yoke Seal 212 0048 Bullnose seal Benelux 303 0030 Cylinder Gauge Boot 303 0060 Swivel Cylinder Gauge 0 3mm Restriction 9 Distributor Details Australia BOC Healthcare Customer Service Centre 10 Julius Avenue North Ryde Sydney NSW 2113 T 61 0 2 8874 4400 F 61 0 2 9886 9000 www boc com au New Zealand BOC Healthcare Customer Service Centre 970 992 Great South Road Penrose 1061 ...

Reviews: