Ultrasound power
The left rotary knob is used to adjust the ultrasound power from 10 %
to 100 % (in 10 % increments).
Duty cycle
The duty cycle is the ratio between ultrasound ON and ultrasound
OFF over an interval of time of 80 milliseconds. Set the duty cycle
using the arrow keys.
The last setting of the duty cycle is kept in memory at power off.
The handpiece and probe error detection is possible only with
a duty cycle set at 100 % (see Troubleshooting chapter).
Activation
• Press the ultrasound (left) foot pedal halfway to activate the
suction (pinch valve opens).
• Press the ultrasound (left) foot pedal completely to activate the
ultrasound handpiece and probe with suction.