Clinician Trial Manual
Clinician Trial Manual
en
3. Refer to “Securing the Trial Lead” in the appropriate Leads Directions for Use.
If desired, secure the OR Cable to the skin using standard dressing technique.
•
OR Cables may be placed side by side and gauze may be used as padding. The area may
be covered with an adhesive and secured with a bandage or dressing.
Fitting the Trial Belt to the Patient
1. Determine which side of the patient is best for placement of the ETS.
2. Fit the trial belt to the patient with the pocket opening facing the patient’s back.
3. Fasten the belt to the patient.
4. If desired, cut off excess length of the trial belt.
5. Place the ETS in the belt pocket with the stimulation ON/OFF button facing up.
6. If OR Cable Extension(s) were used, ensure that they are no longer connected to the OR
Cable(s).
7. Connect the OR Cable(s) to the ETS.
8. Fasten the belt pocket to secure the ETS.
B
A
C
D
1
2
5
3
4
Image is intended for illustrative purposes only.
1 – OR Cable
2 – OR Cable Connection to ETS
3 – ETS
4 – Trial Belt
5 – Lead or Extension
Replacing the External Trial Stimulator Battery
1. Ensure that stimulation is off (the indicator light is not blinking)
before opening the ETS battery compartment.
2. On the rear of the ETS, push in slightly and slide down the bat-
tery compartment cover.
A
B
C
D
Slide
Insert
Batteries
Insert
Batteries
Slide
Summary of Contents for SC-4108
Page 47: ...Testhandleiding voor artsen 92395602 02 43 van 233 nl Deze pagina is met opzet leeg gelaten ...
Page 107: ...Klinikerens prøvehåndbok 92395602 02 103 av 233 no Denne siden er blank med hensikt ...
Page 119: ...Klinikerprøvehåndbog 92395602 02 115 af 233 da Denne side er tilsigtet tom ...
Page 131: ...Kliinikon kokeilukäsikirja 92395602 02 127 233 fi Tämä sivu on tarkoituksellisesti tyhjä ...
Page 237: ...Clinician Trial Manual 92395602 02 233 of 233 This page intentionally left blank ...