To do this, first measure blood pressure on both arms and sub-

sequently always on the arm which produced the higher rea-

ding.

Place cuff in such a manner that the lower cuff edge finishes

ca. 2 – 3 cm above the elbow and the centre of the rubber bag

rests on top of the artery.

Taking the Blood Pressure:

Inflate the cuff to above the systolic pressure. The first strong

pumping action automatically closes the pressure release

valve. By lightly pressing the button of the air release valve the

deflation speed can be finely adjusted. The World Health Or-

ganisation (WHO) recommends a deflation speed of 2 – 3

mmHg per second. To avoid a reading error, read the pressure

values perpendicular to the scale.

When the measurement is completed, firmly press the air re-

lease button once more. This fully opens the release valve to al-

low all the remaining air to escape from the cuff.

Technical Data:

Measurement range: 0 – 300 mmHg

Accuracy of pressure indicated: ± 3 mmHg or 2% of the

reading (whichever is greater)

Storing conditions:

–20°C – 70°C, 15 – 85% rel. humidity

Operating conditions:

10°C – 40°C, 15 – 85% rel. humidity

Cleaning of Unit and Cuff:

– To clean the unit – only use a soft dry cloth.

– To clean the cuff – first remove the rubber bag, the cuff can

then be washed by hand at maximum 30 degrees Celcius.

Disinfection:

For disinfectant wipes (at least 5 minutes exposure time) of the

device and the cuff, we recommend the disinfectant liquid Micro-

zid Liquid Sensitiv (Schülke & Mayr).

To disinfect the cuff, we recommend spray disinfection.

boso_profitest_GB_1802sd.qxd:boso_profitest(GB)_604  06.02.2018  15:17 Uhr  Seite 5

Summary of Contents for Profitest

Page 1: ...Users Instructions boso profitest ...

Page 2: ......

Page 3: ... dealer in accordance with the relevant distributor di rectives A modification of the device and or the accessories is not per mitted This can lead to measurement errors The device can be used by any user skilled in auscultatory blood pressure measurement Purpose Non invasive recording of systolic and diastolic blood pres sure in humans The instrument can be used with any arm cir cumference as sho...

Page 4: ...ldren aged 3 to 12 phase IV K4 pregnant women phase V K5 except when sounds are audible when the cuff is deflated in this case phase IV K4 Korotkoff sounds should be used K5 is the point at which tones heard through the stethoscope are no longer audible K4 is the point at which tones heard through the stethoscope change from a clear sound to a muffled sound Only use the prescribed boso cuffs with ...

Page 5: ...e pressure values perpendicular to the scale When the measurement is completed firmly press the air re lease button once more This fully opens the release valve to al low all the remaining air to escape from the cuff Technical Data Measurement range 0 300 mmHg Accuracy of pressure indicated 3 mmHg or 2 of the reading whichever is greater Storing conditions 20 C 70 C 15 85 rel humidity Operating co...

Page 6: ...d tear e g cuff transport damages and any damage cau sed by improper handling e g non compliance with the in structions for use Damages due to disassembly by unautho rized persons are also excluded from warranty No claims for damages against us are substantiated by the warranty In the case of justified warranty claims the device has to be sent along with the original invoice to BOSCH SOHN GMBH U C...

Page 7: ......

Page 8: ...BOSCH SOHN GmbH u Co KG Bahnhofstraße 64 72417 Jungingen Germany Telephone 49 0 74 77 92 75 0 Fax 49 0 74 77 10 21 Internet www boso de e Mail zentrale boso de 02 2018 0124 ...

Reviews: