Do not inflate to above 300 mmHg!

The actual measurement should not exceed 2 minutes.

An interval of at least 2 minutes has to be allowed between two

measurements.

The patient must follow these basic rules when having his/her

blood pressure measured:

– sit comfortably

– not cross his/her legs

– rest his/her back and arms on the chair or another surface

– have the centre of the cuff placed on the upper arm at the

level of the heart

– relax as much as possible and keep silent while his/her

blood pressure is being measured

– be calm for 5 minutes before the first reading

The following phases are recommended in the case of auscul-

tatory measurement

– adults: phase V (K5)

– children aged 3 to 12: phase IV (K4)

– pregnant women: phase V (K5), except when sounds are

audible when the cuff is deflated: in this case phase IV (K4)

Korotkoff sounds should be used.

K5 is the point at which tones heard through the stethoscope

are no longer audible

K4 is the point at which tones heard through the stethoscope

change from a clear sound to a muffled sound.

Only use the prescribed boso cuffs with their special

connector.

The cuff must be selected such that it fits the arm circumfe-

rence.

Inscriptions:

“mmHg” is the unit in which blood pressure is measured

“shock protected” = the instrument has been tested and found

to work correctly after being dropped from a height of up to one

metre.

Fitting the Cuff:

The cuff must not be fitted over wounds as this could lead to

further injury.

Blood pressure must always be measured on the arm with the

higher pressure values.

boso_profitest_GB_1802sd.qxd:boso_profitest(GB)_604  06.02.2018  15:17 Uhr  Seite 4

Summary of Contents for Profitest

Page 1: ...Users Instructions boso profitest ...

Page 2: ......

Page 3: ... dealer in accordance with the relevant distributor di rectives A modification of the device and or the accessories is not per mitted This can lead to measurement errors The device can be used by any user skilled in auscultatory blood pressure measurement Purpose Non invasive recording of systolic and diastolic blood pres sure in humans The instrument can be used with any arm cir cumference as sho...

Page 4: ...ldren aged 3 to 12 phase IV K4 pregnant women phase V K5 except when sounds are audible when the cuff is deflated in this case phase IV K4 Korotkoff sounds should be used K5 is the point at which tones heard through the stethoscope are no longer audible K4 is the point at which tones heard through the stethoscope change from a clear sound to a muffled sound Only use the prescribed boso cuffs with ...

Page 5: ...e pressure values perpendicular to the scale When the measurement is completed firmly press the air re lease button once more This fully opens the release valve to al low all the remaining air to escape from the cuff Technical Data Measurement range 0 300 mmHg Accuracy of pressure indicated 3 mmHg or 2 of the reading whichever is greater Storing conditions 20 C 70 C 15 85 rel humidity Operating co...

Page 6: ...d tear e g cuff transport damages and any damage cau sed by improper handling e g non compliance with the in structions for use Damages due to disassembly by unautho rized persons are also excluded from warranty No claims for damages against us are substantiated by the warranty In the case of justified warranty claims the device has to be sent along with the original invoice to BOSCH SOHN GMBH U C...

Page 7: ......

Page 8: ...BOSCH SOHN GmbH u Co KG Bahnhofstraße 64 72417 Jungingen Germany Telephone 49 0 74 77 92 75 0 Fax 49 0 74 77 10 21 Internet www boso de e Mail zentrale boso de 02 2018 0124 ...

Reviews: