Instructions for use
Performing a test
3.3 Starting a test
Scanning the sample
Place the
barcode
of the sample underneath the scanner.
A white light beam with a red dot indicates that scanning
is in progress.
The sample data will be
displayed
on the screen.
4
C O N T I N U E
enabled
Alternatively press
edit
to enter the sample data manually.
• Please check
scanned ID
and
displayed ID
to ensure they match.
•
Check
that no other test is running or check the estimated remaining test time
before opening
the cartridge pouch.
• Sample data will be matched automatically when connected to an HIS/LIS.
• Alternatively enter sample data manually.
Scanning the cartridge
Place the
cartridge code (DMC)
underneath the scanner.
A white light beam with a red dot indicates that scanning
is in progress.
The
test description
and further
data
will be displayed
on the screen.
• When scanning the cartridge, the analyser will check the expiry date of the cartridge.
It will only accept cartridges that have not expired.
• If cartridge scanning fails, follow the instructions displayed on the screen.
Inserting the sample
Place
the cartridge on a flat surface.
Insert
the
correct sample type
as described in the
corresponding application-specific instructions for use.
The necessary
volume
is indicated on the cartridge label.
After inserting the sample,
close
the cartridge lid until
it clicks.