1 689 979 642
2014-11-03
|
Robert Bosch GmbH
22 | MMD 302 | Connection and operation
en
4.6.2
Measurement cables can be wrapped
i
Measurement cables can be wrapped around the
appliance and secured in the measurement cable
holders.
4.6.3
Erecting
i
With the protective rubber cover, the appliance can
be erected on a level surface or suspended on a
door or a nail.
4.6.4
Protection
i
The measuring device can also be inserted back to
front in the protective rubber cover in order to pro-
tect the LCD against scratching when not in use.
5. Maintenance
!
The MMD 302 is not to be repaired. The MMD 302 has
no parts of any use to the user. Repairs can only be
performed by qualified customer service personnel.
5.1
Cleaning
!
Always switch off the MMD 302 for cleaning.
The housing and the LCD of the MMD 302 are only
to be cleaned using a soft cloth and neutral cleaning
agents. Do not use abrasive cleaning agents and coarse
workshop cloths. The rubber cover can be removed for
cleaning purposes.
Do not use alcohol-based cleaning agents, as these
could leave residue.
5.2
Spare parts and parts subject to
wear
Designation
Part Number
Cable set, red and black measuring cables
1 687 011 323
Connecting terminal, red
1 681 354 023
Connecting terminal, black
1 681 354 024
Protective rubber cover
1 680 591 026
1A/600V IR 10KA fast active fuse, Buss-
mann BBS-1 or similar rating
1 684 520 023
15A/600V IR 10KA fast active fuse, Buss-
mann KTK 15 or similar rating
1 684 520 024
Rubber seal for housing
1 681 006 031
5.3
LCD difficult to read
If the LCD is difficult to read after the appliance has
been switched on, you should check the batteries as
follows:
Turn the measurement-mode switch (fig. 1, pos. 8) to
the
position, and switch over to diode testing by
means of the blue function button (fig. 1, pos. 2). The
battery voltage is now displayed. The displayed voltage
should be between 2.8 V and 3.2 V in order to guar-
antee safe operation of the MMD 302. The batteries
should be changed if this value lies below 2.5 V.