![Bosch Carephone 61 User Manual Download Page 14](http://html.mh-extra.com/html/bosch/carephone-61/carephone-61_user-manual_42968014.webp)
14
en | Installation
Carephone 61
F.01U.239.30x | V1 | 2011.06
User Manual
Bosch Security Systems
selection enters standby mode (see
Section 6.2.4 Programming
steps, page 17
). The unit can now be programmed.
5.3.2
IP connection
Refer to the user manual of the IP module.
5.3.3
VoIP connection
When using a VoIP connection, be aware of the following
limitations: network availability, priority of the emergency call,
power failures.
5.3.4
Connection configuration
To check whether an emergency call can be sent:
1.
Make a phone call as you normally would.
2.
When the phone call is in progress, press the red button.
If the setup is correct, the phone call will be interrupted.
5.4
Deinstallation
To switch off the Carephone 61 and the backup battery, remove
the power plug from the unit.
NOTICE!
The backup battery will now charge up. The specified standby
time is available when the battery is fully charged, after 16
hours.
NOTICE!
In the event of an emergency, the unit must be able to send out
an emergency call even if a phone call is in progress at the time.
The Carephone 61 must interrupt the phone call.
CAUTION!
With a private branch exchange (PABX) there is no guarantee
that a phone call will be interrupted by the emergency call.
NOTICE!
If you remove the power plug from your power outlet, the unit
announces "
power failure"
. The unit will switch to battery
operation.