Setting the
leakage test interval
9.6
Setting the HC test
9.7
Setting CO
vrai
This menu enables you to define whether
a leakage test must be performed each
time the BEA is switched on or at certain
intervals (in days).
!
Observe legal requirements when
making your entry!
Select the appropriate setting with the
o
and
u
keys and confirm with
E
or by
clicking
Next
.
If you select and confirm the option
Enter
leakage test interval
, proceed to enter
the appropriate interval in days using the
keyboard, e.g.:
7
days.
Confirm your entry with
E
or by clicking
OK
.
The
OK
box is displayed for 1 s.
Entry is now complete.
In this menu you can specify whether or
not an HC residues test should be perfor-
med prior to each measurement.
Select the appropriate setting with the
o
and
u
keys and confirm with
E
or by
clicking
Next
.
The
OK
box is displayed for 1 s.
Entry is now complete.
This is where you can enable the change of
setting for CO
vrai
. This setting work is obli-
gatory as it is a prerequisite of configurati-
on in the
Set parameters
menu (Sec. 4.5).
!
Observe legal requirements when
making your entry!
Select the appropriate setting with the
o
and
u
keys and confirm with
E
or by
clicking
Next
.
The
OK
box is displayed for 1 s.
i
If you select the option CO
vrai
setting
disabled, the CO
vrai
test in the Set
parameters menu (Sec. 4.5.4) can no
longer be enabled.
Summary of Contents for BEA 150
Page 91: ...92 12 Overview 12 1 Electrical wiring Siemens analysis chamber 459716 2Ha ...
Page 94: ...95 12 4 Hosing Siemens analysis chamber X26 X24 X25 X22 X5 LP 223 X19 459716Ht ...
Page 96: ...97 12 6 Electrical wiring Andros analysis chamber 4597162Ht ...
Page 99: ...100 12 9 Hosing Andros analysis chamber 459869_03Ht ...
Page 100: ...101 12 10 Gas circuit diagram Andros analysis chamber ...