57
8.5
Messages about printing functions
No.
Message
Possible cause
Remedy
Section
1200 Internal printer not connected
Loose interconnecting
Check that interconnecting
cables
cables secure,
check continuity
12.1 and 12.4
Faulty controller PCB of
internal printer
Replace controller PCB
10.21.3
Error in control module PCB Replace control module PCB
10.27
1201 Printer not ready
No paper inserted
Insert paper
1202 Other printer errors
1203 Incorrect text to be entered in
printer memory
Software error
Re-boot BEA, if error still
1204 Error linking data module
present:
1205 Error unlinking data
Re-install system software
5.23
1206 Error opening internal printer
RAM error
Check printer driver
5.11 and 5.12
interface
Error in control module PCB Replace control module PCB
10.27
1207 Error opening external printer
interface
8.6
Messages about hardware test functions
No.
Message
Possible cause
Remedy
Section
1300 Time out – no characters received
Error in control module PCB Replace control module PCB
10.27
1301 Error opening serial
Software error
Reboot BEA, if error still
interface
present: Re-install system
software
5.23
1302 Error writing serial interface
Faulty interface
Perform interface test
5.1 - 5.4
1203 Error reading serial interface
Error in control module PCB Replace control module PCB
10.27
1304 Send data different from receive data
Error in control module PCB Replace control module PCB
10.27
1305 Error opening disk interface
Software error
Reboot BEA, if error still
present:
Re-install system software
5.23
Error in control module PCB Replace control module PCB
10.27
1306 Error reading disk
Faulty disk
Insert new disk
1307 Error making temporary
No OS9-formatted disk
Format disk using OS9 Service
see EDIS
filename
inserted
disk.
1308 Error creating temporary file
Error in voltage supply
Check voltage supply
10.2 and 10.3
1309 Error writing on disk
Loose interconnecting
Check that interconnecting
1310 Error reading back
cables
cables secure, check continuity
12.1 and 12.4
1311 Error comparing data
Hard disk error
Perform HD test
5.8
Format HD using Service disk
11.1.3
If error still present:
Replace HD
10.23
Error in control module PCB Replace control module PCB
10.27
1312 Time format error
Software error
Reboot BEA, if error still present:
1313 Error setting time
Re-install system software
5.23
Summary of Contents for BEA 150
Page 91: ...92 12 Overview 12 1 Electrical wiring Siemens analysis chamber 459716 2Ha ...
Page 94: ...95 12 4 Hosing Siemens analysis chamber X26 X24 X25 X22 X5 LP 223 X19 459716Ht ...
Page 96: ...97 12 6 Electrical wiring Andros analysis chamber 4597162Ht ...
Page 99: ...100 12 9 Hosing Andros analysis chamber 459869_03Ht ...
Page 100: ...101 12 10 Gas circuit diagram Andros analysis chamber ...