ACS 553, ACS 563 Service Manual
1 689 975 295
2021-04-13
Robert Bosch GmbH
18
Introduction
Edit Print Header
To make changes to the printer text (available
in Setting menu) appearing in the header on
each printout:
1. Call up the Main Menu.
2. Select
NEXT
icon.
3. Select
MENU
icon.
4. Select
Edit Print Header
in the Settings
Menu. The cursor is in the first field.
5.
Update the text by using the arrows and the
multi-tap interface on the numeric keypad:
• BACKSPACE
icon acts as a backspace
key.
• Right
or
left
arrow moves the cursor to the
right or to the left.
• Zero
(0) key also acts as a spacebar.
• Up
and
Down
arrows navigate between
the rows.
6. Select
OK
icon to save the changes and
return to the Settings Menu.
7. Select
ESC
icon to exit and return to the
Settings Menu.
Replace Printer Paper (only for
variants with printer)
To install a new paper roll in the printer:
1. Remove the cover on the printer by pulling
out on the tab (2) as shown in Figure 1-7.
2. Remove the paper core.
3. Install the new roll of paper with the end of
the paper at the top of the roll.
4. Assemble the cover onto the printer with the
leading edge of the paper over the roller (1).
1
2
Figure 1-7. Replace Print Paper
Calibration Check
This function is used to ensure the machine’s
internal scale is always calibrated. To perform
this check make use of Bosch calibration kit
(optional) including sample weight and magnet.
First time magnet must be installed.
1. Verify the magnet on the bottom of the
machine is clean. Refer to Figure 1-8.
2. Call up the Main Menu.
3. Select
NEXT
icon.
4. Select
MENU
icon.
5. Select
Calibration Check
from the Mainte-
nance Menu. The machine displays:
PLACE CALIBRATION WEIGHT ON MAGNET
LOCATED ON THE BOTTOM OF MACHINE
6. Attach the calibration weight to the magnet
(1) on the bottom of the machine.
7. Select
OK
icon to continue.
ALERT: Unit shown “Reading, do not
disturb”
.
The machine displays:
REMOVE CALIBRATION WEIGHT
FROM MAGNET LOCATED
ON THE BOTTOM OF MACHINE
8.
Remove calibration weight from magnet
located on the bottom of machine.
9. Select
OK
icon to continue.
ALERT: Do not touch the unit in this
phase!
• If the display shows
Calibration passed
the scale is calibration. Select
OK
icon to
return to the Maintenance Menu.
• If the display shows
Calibration
failed
the
scale is out of calibration. Select
RETRY
icon to retry. If calibration continues to fail,
contact an authorized service center for
assistance.
1
Figure 1-8. Magnet Location