
ED01_REV01-2021 IFU OB500_ENG
18- 32
8.1.
Re-use of OB-J FA suction jar
The steps required to separate the jar of secretions from the suction unit, disassemble it and assemble it after cleaning
and disinfection are described below. Before starting, wear protective gloves, also covering your forearms, mouth and
protecting your eyes.
Remove the antibacterial filter from the cover by turning it on its seat and
discard it.
Remove the lid from the jar by pressing lightly on the jar and levering the lid
flap. Empty the contents of the jar.
Remove the overflow valve from the lid.
Separate all the parts.
Parts that make up the lid:
•
Polypropylene cage yellow
•
Polypropylene float, yellow
•
Red silicone gasket
•
Red polypropylene lid
DANGER
Risk of infection due to leakage of potentially contaminated substances during emptying of
secretions. Possible transmission of life threatening infections. Always use suitable PPE and
disinfectants as stipulated by hospital regulations and the relevant authorities.
Beware of certain disinfectants that may stain the jar of secretions and its parts even without
damaging it.
8.2.
Cleaning, disinfection and/or sterilisation of OB-J FA suction jar and silicone tube
The suction jar and the silicone tube can be cleaned with specific non-abrasive substances for cleaning medical devices.
Alcohol-based cleaning agents can be used if diluted appropriately (follow the instructions for use on the label of the
disinfectants). Avoid using coloured disinfectants as they may stain the plastic of the jar and the silicone tube, reducing
its transparency. After disposing of the disposable antibacterial filter and Yankauer suction catheter, complete with
tubing, place the reusable parts in warm water (temperature not exceeding 60°C to avoid scalding) containing a diluted
disinfectant for medical devices. Rinse thoroughly and, if necessary, use a non-abrasive brush to remove any deposits.
After washing, dry all parts. Refer to the cleaning and disinfection plan on the following pages. In the event of serious
contamination,
always refer to the
instructions of medical personnel and the competent authorities. If necessary,
sterilize "REUSABLE PARTS" (see above) with steam autoclaves at a maximum temperature of 121°C for a maximum of
15-20 minutes (typical cycle). Do not use autoclaves with pressures above 2 bar (200 kPa). The jar should be placed
vertically upside down. At the end of the cycle, allow the parts to cool to room temperature and check that they are
undamaged and not deformed.