7.1.
Autoclavable secretion collection jar OB-J FA
The jar is made of transparent plastic (medical-grade polypropylene). It includes
the jar (1), the snap-on lid (2), the anti-reflux valve (3) and the 90° plastic
connection (4). The jar lid allows direct insertion of the antibacterial filter (from
the outside). The autoclavable jar can be sterilised conventionally in a steam
autoclave at a maximum temperature of 121° C and a pressure of 2 bar (200 kPa).
The jar must be replaced if it is deformed, broken or cracked. The suction jar must
be used vertically to prevent the anti-reflux valve from tripping. If the anti-reflux
valve is triggered, switch off the device and disconnect the tube connected to the
suction unit, remove the antibacterial filter to rebalance the pressure inside the
jar.
Service life OB-J FA
The jar of secretions must be replaced after 30 sterilisation cycles or 5 years from the
date of manufacture.
7.2.
Antibacterial filter
The protective filter protects the suction circuit from any contaminants sucked
in during use. The filter is manufactured from hydrophobic PTFE material which
prevents fluids from entering the pneumatic circuit. Working in conjunction with
the overflow valve on the jar, the filter isolates the pneumatic suction pump
from gases and fluids.
The filter is disposable and must be replaced after each
use
. In the event of contamination, discolouration and increased suction
resistance, it must always be replaced. The filter is not produced by the Boscarol
company.
Antibacterial
filter
If the device is used on patients where the infection status is unknown,
always replace the filter
after use on the same patient. This will prevent even serious contamination of the environment
in which the device is placed and therefore of operators and patients. If, on the other hand, it is
known and/or there is no risk of indirect contamination, it is advisable to replace the filter after
each work shift or in any case when the suction level decreases or the filter changes colour.
Risk of infection
•
Never use the device without the antibacterial filter. Please always keep at least three spare
filters in case of emergency.
•
Always wear gloves and personal protective equipment when changing the antibacterial filter
and emptying the suction jars.
•
Before each use, check that the filter is dry and clean (it must not be any colour other than
white). Replace the wet or contaminated filter with a new one.
•
Never reuse the antibacterial filter (disposable).
7.3.
OB-J FM: jar of secretions for SERRES® disposable bags
The OB-J suction jar for SERRES® disposable bags is made of transparent plastic
(medical grade polypropylene). It comprises a container (1), a SERRES® disposable bag
adapter (2), a red 90 degree connector (3) and a SERRES® disposable bag (4). The
antibacterial filter is integrated in the lid of the disposable bag and prevents aspirated
fluids from entering the suction unit. The suction jar can be sterilised in a conventional
steam autoclave at a maximum temperature of 121° C and a pressure of 2 bar (200
kPa). The disposable bag must be replaced after use on the same patient or if it is full.
When used in a domestic environment, the jar of secretions can be cleaned using a
special detergent to ensure disinfection of medical devices. Contact Boscarol for
information on disinfectants.