- 14 -
4 Technical indexes
Meter grade
:
0.01 grade
Measuring range
pH
0-14pH
mV
-1999…1999
Temp
-5
℃
---105
℃
Resolution
pH
0.01pH
mV
1mV
Temp
0.1
℃
Electronic unit
measurement error
pH
±0.01pH
mV
±1mV
Temp
±0.3
℃
pH calibration
1point, 2 point, or 3 point
Isoelectric point
pH 7.00
Buffer solution
8 groups
Power supply
DC6V/20mA
;
4 x AA/LR6 1.5 V or NiMH 1.2 V and
chargable
Size/Weight
230×100×35(mm)/0.4kg
Display
LCD
pH input
BNC
,
resistor >10e+12Ω
Temp input
RCA(Cinch)
,
NTC30kΩ
Data storage
Calibration data
;
198 groups measurement data
(
99
groups for pH
、
mV each)
Working
condition
Temp
5...40
℃
Relative humidity
5%...80%(without condensate)
Installation grade
Ⅱ
Pollution grade
2
Altitude
<=2000m
Summary of Contents for PHS-1701
Page 1: ...Portable pH Meter User Manual...